手持式、基于智能手机的心律记录设备诊断准确性的验证。

Validation of diagnostic accuracy of a handheld, smartphone-based rhythm recording device.

机构信息

Department of Cardiology, Karolinska Institutet and Karolinska University Hospital, Stockholm, Sweden.

出版信息

Expert Rev Med Devices. 2023 Jan;20(1):55-61. doi: 10.1080/17434440.2023.2171290. Epub 2023 Jan 27.

Abstract

AIM

Evaluation of the diagnostic accuracy of a rhythm recording device, for detection of atrial tachyarrhythmia (ATA) and atrial fibrillation (AF) compared to 12-lead-electrocardiogram (12-L-ECG).

RESEARCH DESIGN AND METHODS

Two hundred 12-L-ECGs (reference standard) and Coala Heart Monitor (CHM) recordings (index test) were collected from 189 patients. Two electrophysiologists independently performed manual analysis of all 12-L-ECGs and CHM recordings in random order. The CHM recordings were also analyzed by an automatic algorithm and compared to the results of the reference standard.

RESULTS

Manual analysis of CHM for ATA showed a sensitivity of 98.9% (95% confidence interval (CI): 94.0-100) and a specificity of 100% (CI: 96.6-100). Manual analysis for AF had a sensitivity of 100% (CI: 95.3-100) and a specificity of 97.5% (CI: 93.0-99.5). Automatic analysis for ATA showed a sensitivity of 93.5% (CI: 86.3-97.6) and a specificity of 92.6% (CI: 85.9-96.7). Automatic analysis for AF showed a sensitivity of 97.4% (CI: 91.0-99.7) and a specificity of 86.1% (CI: 78.6-91.7).

CONCLUSION

CHM has a very high accuracy for ATA and AF in manual analysis and a high accuracy for ATA and AF in automatic analysis, making the device suitable for screening.

摘要

目的

评估一种心律记录设备在检测房性心动过速(ATA)和心房颤动(AF)方面的诊断准确性,与 12 导联心电图(12-L-ECG)相比。

研究设计和方法

从 189 名患者中收集了 200 份 12-L-ECG(参考标准)和 Coala Heart Monitor(CHM)记录(指数测试)。两名电生理学家独立地以随机顺序对所有 12-L-ECG 和 CHM 记录进行手动分析。CHM 记录也由自动算法进行分析,并与参考标准的结果进行比较。

结果

CHM 对 ATA 的手动分析显示出 98.9%的敏感性(95%置信区间(CI):94.0-100)和 100%的特异性(CI:96.6-100)。AF 的手动分析敏感性为 100%(CI:95.3-100),特异性为 97.5%(CI:93.0-99.5)。ATA 的自动分析显示出 93.5%的敏感性(CI:86.3-97.6)和 92.6%的特异性(CI:85.9-96.7)。AF 的自动分析显示出 97.4%的敏感性(CI:91.0-99.7)和 86.1%的特异性(CI:78.6-91.7)。

结论

CHM 在手动分析中对 ATA 和 AF 具有非常高的准确性,在自动分析中对 ATA 和 AF 具有较高的准确性,使该设备适合进行筛查。