Bioanalysis. 2022 Dec;14(23):1487-1496. doi: 10.4155/bio-2022-0202. Epub 2023 Jan 27.
An automated method for the measurement of blood tacrolimus on volumetric absorptive microsampling (VAMS) devices was developed. VAMS devices prepared by the automated method were compared with those prepared by the existing manual method (n = 284; mean concentration: 8.0 μg/l; range: 0.6-18.1). The performance of both methods was comparable. Passing-Bablok regression demonstrated an acceptable correlation (y = -0.449 + 1.06x). Bland-Altman analysis demonstrated acceptable agreement (mean bias: -0.007 μg/l; standard deviation: 1.536). Automation reduced operator touch time by 40 min (48-sample batch). Automated preparation of VAMS devices reduced touch time and improved process consistency, facilitating high-throughput testing and transformation of existing laboratory workflows. Automation did not improve precision for VAMS devices but did so for liquid blood samples.
建立了一种用于在体积吸收微采样(VAMS)设备上测量血他克莫司的自动化方法。比较了用自动化方法和现有的手动方法制备的 VAMS 设备(n=284;平均浓度:8.0μg/l;范围:0.6-18.1)。两种方法的性能相当。Passing-Bablok 回归显示出可接受的相关性(y=-0.449+1.06x)。Bland-Altman 分析显示出可接受的一致性(平均偏差:-0.007μg/l;标准偏差:1.536)。自动化操作减少了操作人员的接触时间 40 分钟(48 个样本批)。VAMS 设备的自动化制备减少了接触时间并提高了过程一致性,便于高通量检测和现有实验室工作流程的转化。自动化提高了液体血样的精密度,但没有提高 VAMS 设备的精密度。