Safety and feasibility of the Wahoo Hybrid Access System, a dual-mode guide catheter, in a range of neuroendovascular procedures.

作者信息

Waters Michael J, Vargas Jan, Turk Aquilla, Chaudry Imran, Turner Raymond D

机构信息

Department of Neurosurgery and Neuroendovascular Surgery, 3626Prisma Health Upstate, Greenville, SC, USA.

出版信息

Interv Neuroradiol. 2023 Feb 7:15910199231155033. doi: 10.1177/15910199231155033.

Abstract

BACKGROUND

There is a clinical need for a guide catheter with combined stability and navigability, which can be used in a biaxial system for neuroendovascular procedures in place of triaxial systems.

OBJECTIVE

To assess the safety and feasibility of the Q'Apel Medical Wahoo Hybrid Access System, a dual-mode 0.072″ internal diameter guide catheter, in a range of neuroendovascular procedures.

METHODS

We performed a retrospective analysis of consecutive cases from a high-volume tertiary center in which the Wahoo Hybrid Access System was used as the guide catheter. Characteristics of the patients, vascular lesions, procedure, and procedural complications were assessed.

RESULTS

A total of 102 patients were included for analysis. Vascular lesions were in the anterior circulation in 90 of 102 (88%), and posterior circulation in 12 of 102 (12%). Eighty-four cases were ruptured or unruptured aneurysm embolization procedures, the majority being balloon-assisted coiling (42%) and flow diversion (42%). All cases, including flow diversion, were performed as a biaxial system. There were no instances of prolapse of the catheter beyond the arterial segment in which it was initially placed. The procedure was able to be performed to completion in 101 of 102 (99%) cases. Thromboembolic complications occurred in 5 of 102 (5%); causality in two cases was unrelated to the guide catheter, and three were indeterminate.

CONCLUSIONS

The Wahoo guide catheter is safe and feasible when used in a variety of neuroendovascular procedures. It can accommodate a range of devices, can be safely navigated into distal vasculature, and provides support for a range of procedures, including those which traditionally require triaxial support.

摘要

背景

临床上需要一种兼具稳定性和可操作性的引导导管,可用于双轴系统,以替代三轴系统进行神经血管内介入手术。

目的

评估内径为0.072英寸的双模Q'Apel Medical Wahoo混合接入系统引导导管在一系列神经血管内介入手术中的安全性和可行性。

方法

我们对一家大型三级中心连续使用Wahoo混合接入系统作为引导导管的病例进行了回顾性分析。评估了患者的特征、血管病变、手术过程及手术并发症。

结果

共纳入102例患者进行分析。102例中的90例(88%)血管病变位于前循环,12例(12%)位于后循环。84例为破裂或未破裂动脉瘤栓塞手术,大多数为球囊辅助弹簧圈栓塞(42%)和血流导向(42%)。所有病例,包括血流导向手术,均作为双轴系统进行。没有导管超出其最初放置的动脉节段脱垂的情况。102例中的101例(99%)手术能够完成。102例中有5例(5%)发生血栓栓塞并发症;2例的因果关系与引导导管无关,3例不确定。

结论

Wahoo引导导管在用于各种神经血管内介入手术时是安全可行的。它可以容纳多种器械,能够安全地进入远端血管,并为一系列手术提供支持,包括那些传统上需要三轴支持的手术。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验