Effects of slice selection and diffusion on T2 measurement.

作者信息

Wong S T, Roos M S

机构信息

Lawrence Berkeley Laboratory, University of California, Berkeley 94720.

出版信息

Magn Reson Med. 1987 Oct;5(4):358-65. doi: 10.1002/mrm.1910050407.

Abstract

This study examines slice selection and diffusion as sources of systematic error in T2 estimates obtained from imaging experiments. The selective refocusing pulses in a multislice, two-echo, spin-warp experiment cause slice profiles of sequential echoes to change in shape. This results in bias in T2 images calculated from the two-echo data set. A method to correct the bias is presented. The effect of diffusion on T2 images obtained with typical two-echo imaging sequences was found to be negligible.