An 81 year-old man presented with an asymptomatic juxtrarenal abdominal aortic aneurysm and was subsequently treated with a fenestrated endovascular Anaconda stent-graft. Surveillance imaging within the first postoperative year demonstrated a lower proximal sealing ring fracture. In the second postoperative surveillance year, the upper proximal sealing ring was also fractured with extension of the wire into the right paravertebral space. Despite these sealing ring fractures, there were no endoleak nor visceral stent complications and the patient continued on standard surveillance protocols. There are an increasing number of reports of fractured proximal sealing rings with the fenestrated Anaconda platform. Those analysing the surveillance scans of patients treated with this device should stay vigilant for the development of this complication.
一位 81 岁男性因无症状的肾下腹部主动脉瘤就诊,随后接受了 fenestrated 血管内 Anaconda 支架移植物治疗。术后第一年的监测影像显示较低的近端密封环骨折。在术后第二年的监测中,上近端密封环也发生了骨折,金属丝延伸至右侧椎旁间隙。尽管存在这些密封环骨折,但没有内漏或内脏支架并发症,患者继续按照标准监测方案进行治疗。越来越多的关于 fenestrated Anaconda 平台近端密封环断裂的报告。那些分析使用该装置治疗的患者的监测扫描的人应该对这种并发症的发展保持警惕。