The optical measurement of cerebral oxygen metabolism was evaluated.
Compare optically derived cerebral signals to the electroencephalographic bispectral index (BIS) sensors to monitor propofol-induced anesthesia during surgery.
Relative cerebral metabolic rate of oxygen ( ) and blood flow (rCBF) were measured by time-resolved and diffuse correlation spectroscopies. Changes were tested against the relative BIS (rBIS) ones. The synchronism in the changes was also assessed by the R-Pearson correlation.
In 23 measurements, optically derived signals showed significant changes in agreement with rBIS: during propofol induction, rBIS decreased by 67% [interquartile ranges (IQR) 62% to 71%], by 33% (IQR 18% to 46%), and rCBF by 28% (IQR 10% to 37%). During recovery, a significant increase was observed for rBIS (48%, IQR 38% to 55%), (29%, IQR 17% to 39%), and rCBF (30%, IQR 10% to 44%). The significance and direction of the changes subject-by-subject were tested: the coupling between the rBIS, , and rCBF was witnessed in the majority of the cases (14/18 and 12/18 for rCBF and 19/21 and 13/18 for in the initial and final part, respectively). These changes were also correlated in time ( to , ).
Optics can reliably monitor in such conditions.
对脑氧代谢的光学测量进行了评估。
比较光学衍生的脑信号与脑电图双谱指数(BIS)传感器,以监测手术期间丙泊酚诱导的麻醉。
通过时间分辨光谱和扩散相关光谱测量相对脑氧代谢率( )和血流(rCBF)。将变化与相对BIS(rBIS)的变化进行比较。还通过R-皮尔逊相关性评估变化的同步性。
在23次测量中,光学衍生信号显示出与rBIS一致的显著变化:在丙泊酚诱导期间,rBIS下降了67%[四分位间距(IQR)为62%至71%], 下降了33%(IQR为18%至46%),rCBF下降了28%(IQR为10%至37%)。在恢复期间,观察到rBIS(48%,IQR为38%至55%)、 (29%,IQR为17%至39%)和rCBF(30%,IQR为10%至44%)有显著增加。对每个受试者变化的显著性和方向进行了测试:在大多数情况下(分别在初始和最终部分,rCBF为14/18和12/18, 为19/21和13/18)观察到rBIS、 和rCBF之间的耦合。这些变化在时间上也具有相关性( 与 , )。