Department of Radiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
Magn Reson Med. 2023 Aug;90(2):624-632. doi: 10.1002/mrm.29670. Epub 2023 Apr 24.
To investigate the influence of the sodium (Na) reference tube location in a birdcage coil on the quantification of Na in the calf muscle. Two correction methods were also evaluated.
Eight (4 × 20 mM, 4 × 30 mM Na) reference tubes were placed along the inner surface of the coil and one (30 mM Na) tube more centrally near the tibia. In two volunteers, four repeated UTE scans were acquired. In six calf muscles, the Na concentration was calculated based on each reference tube. Flip angle mapping of a homogenous Na phantom was used for correcting intensity values. Alternatively, a normalized intensity map was used for correcting the in vivo signal intensities. Results were given as range or SD of Na concentration measurements over the reference tubes.
For calf Na measurements, there was limited space for positioning reference tubes away from coil B inhomogeneity. In both volunteers, the Na quantification depended greatly on the reference tube used with a range of up to 10 mM. The central tube location gave a Na quantification close to the mean of the other tubes. The flip angle and normalized signal intensity phantom-based correction methods decreased the quantification variation from 14.9% to 5.0% and 10.4% to 2.7%, respectively. Both correction methods had little influence (< 2.3%) on quantification based on the central tube.
Despite use of a birdcage coil, location of the reference tube had a great impact on Na quantification in the calf muscles. Although both correction methods did reduce this variation, placing the reference tube more centrally was found to give the most reliable results.
研究鸟笼线圈中钠(Na)参比管位置对小腿肌肉中 Na 定量的影响。还评估了两种校正方法。
将 8 个(4×20mM,4×30mM Na)参比管沿线圈内表面放置,一个(30mM Na)管更靠近胫骨放置在中心附近。在两名志愿者中,采集了 4 次重复 UTE 扫描。在 6 块小腿肌肉中,根据每个参比管计算了 Na 浓度。使用均匀 Na 体模的翻转角映射来校正强度值。或者,使用归一化强度图校正体内信号强度。结果以参比管上 Na 浓度测量的范围或 SD 表示。
对于小腿 Na 测量,参比管远离线圈不均匀性的位置空间有限。在两名志愿者中,Na 定量极大地取决于所使用的参比管,范围高达 10mM。中央管位置给出的 Na 定量接近其他管的平均值。翻转角和归一化信号强度基于体模的校正方法将定量变化从 14.9%降低至 5.0%和 10.4%至 2.7%。这两种校正方法对基于中央管的定量影响都很小(<2.3%)。
尽管使用了鸟笼线圈,但参比管的位置对小腿肌肉中 Na 的定量有很大影响。尽管这两种校正方法确实减少了这种变化,但发现将参比管放置在中心附近会产生更可靠的结果。