Gruson Damien, Siham Akdim, Fillée Catherine
Deportment of Clinical Biochemistry, Cliniques Universitaires St-Luc and Universite Catholique de Louvain, Brussels, Belgium.
Pôle de recherche en Endocrinologie, Diabète et Nutrition, Institut de Recherche Experimentale et Clinique, Cliniques Universitaires St-Luc and Université Catholique de Louvain, Brussels, Belgium.
EJIFCC. 2023 Apr 18;34(1):4-9. eCollection 2023 Apr.
The added value of Anti-Müllerian hormone (AMH) measurement is recognized for several clinical applications such as assessment of the ovarian reserve, monitoring of fertilization protocol or in the field of oncofertility. Our study objective was to determine the performances of a novel fully automated chemiluminescent assay for AMH testing.
We evaluated the performances of the Maglumi 800 AMH chemiluminescent immunoassay that applies N-(4-Aminobutyl)-N-ethylisoluminol (ABEI) labels. Assay imprecision was assessed with two levels of control materials. Method comparison was performed with an ultrasensitive AMH ELISA assay (Ansh Laboratories, Inc, Webster, TX, USA) with 88 patients' samples.
The within-run and between-run coefficients of variation (CVs) were below for both low and high internal quality controls. The automated and ELISA methods were significantly correlated. Bland-Altman plot evidenced a bias between the methods with a mean bias of 0.6 ng/mL.
Our preliminary evaluation showed overall good analytical performances for the Maglumi AMH fully automated immunoassay and good concordance with a routinely used assay.