A new continuous noninvasive finger cuff device (Vitalstream) for cardiac output that communicates wirelessly via bluetooth or Wi-Fi.

机构信息

Cooper University Hospital, Camden, NJ, USA.

Caretaker Medical, Charlottesville, VA, USA.

出版信息

BMC Anesthesiol. 2023 May 25;23(1):180. doi: 10.1186/s12871-023-02114-z.

Abstract

BACKGROUND

The new noninvasive Vitalstream (VS) continuous physiological monitor (Caretaker Medical LLC, Charlottesville, Virginia), allows continuous cardiac output by a low pump-inflated, finger cuff that pneumatically couples arterial pulsations via a pressure line to a pressure sensor for detection and analysis. Physiological data are communicated wirelessly to a tablet-based user interface via Bluetooth or Wi-Fi. We evaluated its performance against thermodilution cardiac output in patients undergoing cardiac surgery.

METHODS

We compared the agreement between thermodilution cardiac output to that obtained by the continuous noninvasive system during cardiac surgery pre and post-cardiac bypass. Thermodilution cardiac output was performed routinely when clinically indicated by an iced saline cold injectate system. All comparisons between VS and TD/CCO data were post-processed. In order to match the VS CO readings to the averaged discrete TD bolus data, the averaged CO readings of the ten seconds of VS CO data points prior to a sequence of TD bolus injections was matched. Time alignment was based on the medical record time and the VS time-stamped data points. The accuracy against reference TD measurements was assessed via Bland-Altman analysis of the CO values and standard concordance analysis of the ΔCO values (with a 15% exclusion zone).

RESULTS

Analysis of the data compared the accuracy of the matched measurement pairs of VS and TD/CCO VS absolute CO values with and without initial calibration to the discrete TD CO values, as well as the trending ability, i.e., ΔCO values of the VS physiological monitor compared to those of the reference. The results were comparable with other non-invasive as well as invasive technologies and Bland-Altman analyses showed high agreement between devices in a diverse patient population. The results are significant regarding the goal of expanding access to effective, wireless and readily implemented fluid management monitoring tools to hospital sections previously not covered because of the limitations of traditional technologies.

CONCLUSION

This study demonstrated that the agreement between the VS CO and TD CO was clinically acceptable with a percent error (PE) of 34.5 to 38% with and without external calibration. The threshold for an acceptable agreement between the VS and TD was considered to be below 40% which is below the threshold recommended by others.

摘要

背景

新型非侵入式 Vitalstream(VS)连续生理监测仪(Caretaker Medical LLC,弗吉尼亚州夏洛茨维尔)通过低压充气的手指袖带实现连续心输出量监测,袖带通过压力管与压力传感器气动耦合,以检测和分析动脉搏动。生理数据通过蓝牙或 Wi-Fi 无线传输到基于平板电脑的用户界面。我们评估了该监测仪在心脏手术患者中的性能,与热稀释心输出量进行了比较。

方法

我们比较了心脏手术前和体外循环后连续非侵入式系统与热稀释心输出量的一致性。当临床需要时,通过冰冷盐水冷注射系统常规进行热稀释心输出量测量。所有 VS 与 TD/CCO 数据的比较均为后处理。为了将 VS CO 读数与离散 TD 推注数据相匹配,将 VS CO 数据点前 10 秒的平均 CO 读数与一系列 TD 推注的平均 CO 读数相匹配。时间对准基于医疗记录时间和 VS 时间标记数据点。通过 CO 值的 Bland-Altman 分析和 ΔCO 值的标准一致性分析(排除 15%的区域)评估了与参考 TD 测量值的准确性。

结果

数据分析比较了 VS 和 TD/CCO 的匹配测量对的准确性,包括 VS 绝对值 CO 值和无初始校准的 VS 绝对值 CO 值与离散 TD CO 值的一致性,以及 VS 生理监测仪与参考值的 ΔCO 值的趋势能力。结果与其他非侵入式和侵入式技术相当, Bland-Altman 分析显示在不同患者人群中,设备之间具有高度一致性。鉴于将有效、无线且易于实施的液体管理监测工具扩展到以前因传统技术限制而未覆盖的医院科室的目标,这些结果具有重要意义。

结论

本研究表明,VS CO 与 TD CO 之间的一致性在临床可接受范围内,无外部校准和有外部校准的百分比误差(PE)分别为 34.5%至 38%。VS 与 TD 之间可接受的一致性阈值被认为低于 40%,低于其他人推荐的阈值。