Suppr超能文献

生物可吸收干扰螺钉导致大量骨溶解引起的初次全膝关节置换早期失败

Early Failure of Primary Total Knee Arthroplasty Due to Massive Osteolysis Caused by Bio-Absorbable Interference Screws.

作者信息

Galiveeti Mukund Reddy V, El-Abed Khaldoun, Ahmad Riaz

机构信息

Trauma and Orthopaedics, University Hospitals Bristol and Weston, NHS Foundation Trust, Weston-super-Mare, GBR.

出版信息

Cureus. 2023 Apr 26;15(4):e38143. doi: 10.7759/cureus.38143. eCollection 2023 Apr.

Abstract

We report an unusual case in which an unabsorbed bio-absorbable screw in the tibial tunnel of anterior cruciate ligament reconstruction (ACLR) performed 11 years ago caused massive osteolysis and subsequent failure of total knee arthroplasty (TKA). ACLR was performed using suspensory fixation on the femoral side and a bio-absorbable interference screw on the tibial side. Fragmentation of the bio-absorbable screw at the time of tibial component implantation is thought to have evoked an accelerated inflammatory response, causing osteolysis, which finally resulted in early failure of the TKA.

摘要

我们报告了一例罕见病例,11年前进行的前交叉韧带重建术(ACLR)中,胫骨隧道内未吸收的生物可吸收螺钉导致了大量骨质溶解,随后全膝关节置换术(TKA)失败。ACLR在股骨侧采用悬吊固定,在胫骨侧采用生物可吸收加压螺钉。据认为,在植入胫骨部件时生物可吸收螺钉碎裂引发了加速的炎症反应,导致骨质溶解,最终导致TKA早期失败。