手持式测力计评估翻修全膝关节置换术后患者膝关节肌力的可靠性、同时效度和最小可检测变化。

Reliability, Concurrent Validity, and Minimal Detectable Change of the Hand-Held Dynamometry for the Assessment of Knee Muscle Strength in Patients With Revision Total Knee Arthroplasty.

机构信息

Department of Physiotherapy and Rehabilitation, Faculty of Health Sciences, Erzurum Technical University, Yakutiye-Erzurum, Turkey.

Faculty of Physical Therapy and Rehabilitation, Dokuz Eylul University, Balçova-Izmir, Turkey.

出版信息

Arch Phys Med Rehabil. 2024 Jan;105(1):34-39. doi: 10.1016/j.apmr.2023.05.002. Epub 2023 May 30.

Abstract

OBJECTIVE

To determine the concurrent validity, reliability, and minimal detectable change (MDC) of the hand-held dynamometry (HHD) for knee strength measurement in patients with revision total knee arthroplasty (r-TKA).

DESIGN

A reliability and validity analysis.

SETTING

Orthopedic and physical therapy services of university hospital.

PARTICIPANTS

The study included 42 patients with r-TKA (N=42).

INTERVENTIONS

Not applicable.

MAIN OUTCOME MEASURES

Knee muscle strength assessments were performed by 2 physiotherapists in 3 different sessions by using HHD. Participants were instructed to exert a maximal force for lasting 5 seconds against HHD. The first examiner performed the strength testing, and after 30-minutes rest, the second examiner performed the same procedure for inter-examiner reliability. One hour after the initial testing, the first examiner reperformed the strength testing for intra-examiner reliability. The correlations of the knee extensors and knee flexors strength with 50-foot walking test and 30-second chair stand test were assessed for concurrent validity.

RESULTS

The inter-examiner reliabilities of knee extensors and flexors strength measurements were 0.97 and 0.95, respectively. The SEM and the minimal detectable changes at 95% confidence level (MDC) for knee extensors were 10.39 and 28.65 Newton-meters (Nm), and SEM and MDC for knee flexors were 8.70 and 23.99 Nm, respectively. The intra-examiner reliabilities of knee extensors and flexors strength measurements were 0.96. SEM and MDC for knee extensors were 12.00 and 33.09 Nm, and SEM and MDC for knee flexors were 7.78 and 21.45 Nm, respectively. The knee muscle strength showed strong significant correlations with physical performance tests (all, P<.05).

CONCLUSIONS

The HHD is a reliable and valid method for assessment of static knee strength after r-TKA. The HHD can be used to quantify changes in knee strength and also assists the clinicians to determine the effect of rehabilitation programs on muscle strength following r-TKA surgery.

摘要

目的

确定手持式测力计(HHD)用于测量翻修全膝关节置换术(r-TKA)后膝关节强度的同时效度、信度和最小可检测变化(MDC)。

设计

可靠性和有效性分析。

设置

大学医院的骨科和物理治疗服务。

参与者

该研究纳入了 42 例 r-TKA 患者(N=42)。

干预措施

不适用。

主要观察指标

由 2 名物理治疗师在 3 个不同时段使用 HHD 进行膝关节肌肉力量评估。指导患者对 HHD 施加最大力,持续 5 秒。第一检查者进行力量测试,在 30 分钟休息后,第二检查者进行相同的程序以评估检查者间信度。初始测试后 1 小时,第一检查者重复进行力量测试以评估检查者内信度。评估膝关节伸肌和屈肌力量与 50 英尺步行测试和 30 秒坐站测试的相关性,以评估同时效度。

结果

膝关节伸肌和屈肌力量测量的检查者间信度分别为 0.97 和 0.95。膝关节伸肌的 SEM 和 95%置信水平的最小可检测变化(MDC)分别为 10.39 和 28.65 牛顿米(Nm),膝关节屈肌的 SEM 和 MDC 分别为 8.70 和 23.99 Nm。膝关节伸肌和屈肌力量测量的检查者内信度分别为 0.96。膝关节伸肌的 SEM 和 MDC 分别为 12.00 和 33.09 Nm,膝关节屈肌的 SEM 和 MDC 分别为 7.78 和 21.45 Nm。膝关节肌肉力量与身体表现测试呈显著相关(均 P<.05)。

结论

HHD 是一种可靠且有效的方法,可用于评估 r-TKA 后膝关节的静态力量。HHD 可用于量化膝关节力量的变化,并协助临床医生确定 r-TKA 手术后康复计划对肌肉力量的影响。