Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas (UNICAMP), Piracicaba, São Paulo, Brazil.
Dentomaxillofac Radiol. 2023 Oct;52(7):20230174. doi: 10.1259/dmfr.20230174. Epub 2023 Sep 4.
To subjectively assess radiographs obtained with photostimulable phosphor (PSP) plates exposed to clinical levels of ambient light prior to read-out to potentially set a safe limit for acceptable image quality.
Six dental regions of a dry human skull were X-rayed using PSP plates from VistaScan and Express under four exposure times: 0.1, 0.2, 0.32, and 0.4 s. Before read-out, the PSP plates were exposed to ambient light for 0, 5, 10, 30, 60, and 90 s. Six observers were asked to classify the 288 resulting radiographs as acceptable or unacceptable based on the identification of anatomical structures and global image quality. The number of answers classifying radiographs as unacceptable was used to calculate a rejection rate; a pairwise comparison for better image quality was further conducted among radiographs considered acceptable. Reproducibility was tested by having 25% of all experimental groups reassessed.
Intra- and interobserver agreement ranged from 0.87 to 1.00 and from 0.81 to 0.92, respectively. Exposure of PSP plates to ambient light increased rejection rates mostly as of 10 s. In the pairwise comparison, subtle differences were observed between radiographs obtained with PSP plates not exposed and those exposed to ambient light for 5 s.
Ambient light exposure of PSP plates impairs the image quality of radiographs. A safe limit of ambient light exposure of 5 s for VistaScan and Express should be considered. Ambient light exposure of PSP plates within safe limits can avoid retakes and reduce unnecessary patient exposure to X-rays.
通过对曝光前在临床环境光线下暴露的光激励磷(PSP)板获得的射线照片进行主观评估,为可接受的图像质量设定安全限制。
使用 VistaScan 和 Express 的 PSP 板对干燥人颅骨的六个牙科区域进行 X 射线照射,曝光时间分别为 0.1、0.2、0.32 和 0.4 s。在读取之前,将 PSP 板暴露在环境光下 0、5、10、30、60 和 90 s。六名观察者根据解剖结构和整体图像质量的识别,将 288 张射线照片分为可接受或不可接受。不可接受射线照片的数量用于计算拒绝率;对可接受的射线照片进行进一步的两两比较,以评估更好的图像质量。通过重新评估所有实验组的 25%来测试可重复性。
观察者内和观察者间的一致性分别为 0.87 至 1.00 和 0.81 至 0.92。PSP 板暴露在环境光下会增加拒绝率,尤其是在 10 s 之后。在两两比较中,观察到未暴露和暴露于环境光 5 s 的 PSP 板获得的射线照片之间存在细微差异。
PSP 板暴露在环境光下会降低射线照片的图像质量。应考虑 VistaScan 和 Express 的 PSP 板环境光暴露 5 s 的安全限制。在安全限制内,PSP 板暴露在环境光下可以避免重拍,并减少患者不必要的 X 射线暴露。