荧光板的环境光暴露:是否存在可接受图像质量的安全限值?

Ambient light exposure of photostimulable phosphor plates: is there a safe limit for acceptable image quality?

机构信息

Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas (UNICAMP), Piracicaba, São Paulo, Brazil.

出版信息

Dentomaxillofac Radiol. 2023 Oct;52(7):20230174. doi: 10.1259/dmfr.20230174. Epub 2023 Sep 4.

Abstract

OBJECTIVES

To subjectively assess radiographs obtained with photostimulable phosphor (PSP) plates exposed to clinical levels of ambient light prior to read-out to potentially set a safe limit for acceptable image quality.

METHODS AND MATERIALS

Six dental regions of a dry human skull were X-rayed using PSP plates from VistaScan and Express under four exposure times: 0.1, 0.2, 0.32, and 0.4 s. Before read-out, the PSP plates were exposed to ambient light for 0, 5, 10, 30, 60, and 90 s. Six observers were asked to classify the 288 resulting radiographs as acceptable or unacceptable based on the identification of anatomical structures and global image quality. The number of answers classifying radiographs as unacceptable was used to calculate a rejection rate; a pairwise comparison for better image quality was further conducted among radiographs considered acceptable. Reproducibility was tested by having 25% of all experimental groups reassessed.

RESULTS

Intra- and interobserver agreement ranged from 0.87 to 1.00 and from 0.81 to 0.92, respectively. Exposure of PSP plates to ambient light increased rejection rates mostly as of 10 s. In the pairwise comparison, subtle differences were observed between radiographs obtained with PSP plates not exposed and those exposed to ambient light for 5 s.

CONCLUSIONS

Ambient light exposure of PSP plates impairs the image quality of radiographs. A safe limit of ambient light exposure of 5 s for VistaScan and Express should be considered. Ambient light exposure of PSP plates within safe limits can avoid retakes and reduce unnecessary patient exposure to X-rays.

摘要

目的

通过对曝光前在临床环境光线下暴露的光激励磷(PSP)板获得的射线照片进行主观评估,为可接受的图像质量设定安全限制。

方法和材料

使用 VistaScan 和 Express 的 PSP 板对干燥人颅骨的六个牙科区域进行 X 射线照射,曝光时间分别为 0.1、0.2、0.32 和 0.4 s。在读取之前,将 PSP 板暴露在环境光下 0、5、10、30、60 和 90 s。六名观察者根据解剖结构和整体图像质量的识别,将 288 张射线照片分为可接受或不可接受。不可接受射线照片的数量用于计算拒绝率;对可接受的射线照片进行进一步的两两比较,以评估更好的图像质量。通过重新评估所有实验组的 25%来测试可重复性。

结果

观察者内和观察者间的一致性分别为 0.87 至 1.00 和 0.81 至 0.92。PSP 板暴露在环境光下会增加拒绝率,尤其是在 10 s 之后。在两两比较中,观察到未暴露和暴露于环境光 5 s 的 PSP 板获得的射线照片之间存在细微差异。

结论

PSP 板暴露在环境光下会降低射线照片的图像质量。应考虑 VistaScan 和 Express 的 PSP 板环境光暴露 5 s 的安全限制。在安全限制内,PSP 板暴露在环境光下可以避免重拍,并减少患者不必要的 X 射线暴露。

相似文献

1
Ambient light exposure of photostimulable phosphor plates: is there a safe limit for acceptable image quality?
Dentomaxillofac Radiol. 2023 Oct;52(7):20230174. doi: 10.1259/dmfr.20230174. Epub 2023 Sep 4.
2
Can Digital Enhancement Restore the Image Quality of Phosphor Plate-Based Radiographs Partially Damaged by Ambient Light?
J Imaging Inform Med. 2024 Feb;37(1):145-150. doi: 10.1007/s10278-023-00922-w. Epub 2024 Jan 12.
4
Revisiting dynamic range and image enhancement ability of contemporary digital radiographic systems-part 2: a subjective assessment.
Dentomaxillofac Radiol. 2023 Apr;52(4):20220370. doi: 10.1259/dmfr.20220370. Epub 2023 Mar 29.
6
Disinfection of intraoral photostimulable phosphor plates with 70% ethyl alcohol and 70% isopropyl alcohol alters image contrast and produces image artifacts.
Oral Surg Oral Med Oral Pathol Oral Radiol. 2023 Dec;136(6):741-752. doi: 10.1016/j.oooo.2023.06.011. Epub 2023 Jun 23.
7
Evaluation of the microstructure, chemical composition, and image quality of different PSP receptors.
Braz Oral Res. 2022 Nov 11;36:e130. doi: 10.1590/1807-3107bor-2022.vol36.0130. eCollection 2022.
9
10
Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings.
Braz Oral Res. 2006 Jul-Sep;20(3):207-13. doi: 10.1590/s1806-83242006000300005.

引用本文的文献

1
Influence of a portable X-ray device in the diagnosis of proximal caries lesions.
Braz Oral Res. 2025 May 23;39:e028. doi: 10.1590/1807-3107bor-2025.vol39.028. eCollection 2025.
2
Intraoral digital radiography: A comprehensive report on the technical specifications of current and historical systems.
Imaging Sci Dent. 2025 Mar;55(1):72-89. doi: 10.5624/isd.20240248. Epub 2025 Mar 10.
3
Comparison of objective radiograph quality between radiographs obtained with wall-mounted and handheld X-ray devices.
Imaging Sci Dent. 2025 Mar;55(1):22-27. doi: 10.5624/isd.20240112. Epub 2024 Dec 6.

本文引用的文献

1
Revisiting dynamic range and image enhancement ability of contemporary digital radiographic systems-part 2: a subjective assessment.
Dentomaxillofac Radiol. 2023 Apr;52(4):20220370. doi: 10.1259/dmfr.20220370. Epub 2023 Mar 29.
3
Evaluation of the microstructure, chemical composition, and image quality of different PSP receptors.
Braz Oral Res. 2022 Nov 11;36:e130. doi: 10.1590/1807-3107bor-2022.vol36.0130. eCollection 2022.
4
The importance of the ANSI ADA Standard for digital intraoral radiographic systems-a pragmatic approach to quality assurance.
Oral Surg Oral Med Oral Pathol Oral Radiol. 2023 Jan;135(1):117-128. doi: 10.1016/j.oooo.2022.07.011. Epub 2022 Jul 24.
5
Revisiting dynamic range and image enhancement ability of contemporary digital radiographic systems.
Dentomaxillofac Radiol. 2022 May 1;51(4):20210404. doi: 10.1259/dmfr.20210404. Epub 2021 Dec 14.
6
Can Dental Office Lighting Intensity Conditions Influence the Accuracy of Intraoral Scanning?
Scanning. 2021 May 27;2021:9980590. doi: 10.1155/2021/9980590. eCollection 2021.
7
ALADAIP, beyond ALARA and towards personalized optimization for paediatric cone-beam CT.
Int J Paediatr Dent. 2021 Sep;31(5):676-678. doi: 10.1111/ipd.12797. Epub 2021 May 20.
9
Determination and classification of intraoral phosphor storage plate artifacts and errors.
Imaging Sci Dent. 2019 Sep;49(3):219-228. doi: 10.5624/isd.2019.49.3.219. Epub 2019 Sep 24.
10
A pilot study of the effects of a vacuum sealing barrier method for photostimulable phosphor plates regarding plate damage and prevention of microbiologic cross-contamination.
Oral Surg Oral Med Oral Pathol Oral Radiol. 2018 Sep;126(3):283-290. doi: 10.1016/j.oooo.2018.05.015. Epub 2018 May 29.