Newman Christine, Hartnell Sara, Wilinska Malgorzata, Alwan Heba, Hovorka Roman
Wolfson Diabetes and Endocrinology, Addenbrooke's Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, UK.
Wellcome-MRC Institute of Metabolic Science-Metabolic Research Laboratories, University of Cambridge, Cambridge, UK.
J Diabetes Sci Technol. 2025 Jan;19(1):165-168. doi: 10.1177/19322968231187915. Epub 2023 Jul 28.
We evaluated the performance of the interoperable Cambridge hybrid closed-loop app with FreeStyle Libre 3 glucose sensor, and YpsoPump insulin pump in a real-world setting. Data from 100 users (63 adults [mean ± SD age 41.9 ± 14.0 years], 15 children [8.6 ± 5.2 years)] and 22 users of unreported age) for a period of 28 days were analyzed. Time in range (3.91- 10.0mmol/L) was 72.6 ± 11.1% overall. Time below range (<3.9mmol/L) was 3.1% (1.4-5.1) (median [interquartile range]). Auto-mode was active for 95.8% (91.8-97.9) of time. This real-world analysis suggests that the performance of Cambridge hybrid closed-loop app with this glucose sensor is comparable to other commercially available hybrid closed-loop systems.
我们在实际环境中评估了可互操作的剑桥混合闭环应用程序与FreeStyle Libre 3葡萄糖传感器以及YpsoPump胰岛素泵的性能。分析了100名用户(63名成年人[平均±标准差年龄41.9±14.0岁]、15名儿童[8.6±5.2岁]以及22名未报告年龄的用户)在28天内的数据。总体而言,血糖处于目标范围(3.91-10.0mmol/L)的时间为72.6±11.1%。血糖低于目标范围(<3.9mmol/L)的时间为3.1%(1.4-5.1)(中位数[四分位间距])。自动模式开启时间占95.8%(91.8-97.9)。这项实际环境分析表明,该葡萄糖传感器搭配的剑桥混合闭环应用程序的性能与其他市售混合闭环系统相当。