Validation study of new left atrial appendage closure device measurement by two-dimensional and three-dimensional transesophageal echocardiography.

机构信息

Department of Cardiology, Kurashiki Central Hospital, 1-1-1 Miwa, Kurashiki, 710-0052, Japan.

出版信息

J Echocardiogr. 2024 Mar;22(1):16-24. doi: 10.1007/s12574-023-00619-7. Epub 2023 Aug 29.

Abstract

BACKGROUND

In left atrial appendage closure using WATCHMAN FLX, accurate device measurement by transesophageal echocardiography (TEE) is important. We aimed to experimentally validate appropriate methods of device measurement with two-dimensional (2D) and three-dimensional (3D) TEE compared with actual size.

METHODS

We prepared a full range of device sizes (20, 24, 27, 31, 35 mm), each with five different compression rates. Each device was measured by 2D and 3D TEE at depths of 2, 4, and 6 cm in vitro using inner, outer, and middle line methods. We compared the difference between the actual size by caliper and measurements at each compression rate and depth by the three methods in 2D and 3D TEE.

RESULTS

A total of 450 patterns of measurements were analyzed. The differences using the middle line method were much less than those using the inner and outer line methods in 2D and 3D TEE (2D TEE: 0.45 ± 0.36 vs. 2.55 ± 0.99 vs. 2.59 ± 0.72 mm, p < 0.01; 3D TEE: 0.34 ± 0.27 vs. 2.38 ± 0.69 vs. 1.86 ± 0.77 mm, p < 0.01). Moreover, the differences in measurements by 3D TEE were more accurate than those of 2D TEE in the inner (2.47 ± 1.86 vs. 1.86 ± 0.77 mm, p < 0.01) and middle (0.58 ± 0.37 vs. 0.34 ± 0.27 mm, p < 0.01) line methods.

CONCLUSIONS

Middle line method by 3D TEE is the most reliable approach for device measurement at left atrial appendage closure using WATCHMAN FLX device.

摘要

背景

在使用 WATCHMAN FLX 进行左心耳封堵时,经食管超声心动图(TEE)的准确器械测量非常重要。我们旨在通过实验验证二维(2D)和三维(3D)TEE 与实际尺寸相比,器械测量的合适方法。

方法

我们准备了全系列器械尺寸(20、24、27、31、35 毫米),每个尺寸有 5 种不同的压缩率。在体外,每个器械在 2、4 和 6 厘米的深度通过 2D 和 3D TEE 使用内线、外线和中线方法进行测量。我们比较了在每个压缩率和深度下通过三种方法进行的 2D 和 3D TEE 中的实际尺寸与测量值之间的差异。

结果

共分析了 450 种测量模式。在 2D 和 3D TEE 中,中线方法的差异明显小于内线和外线方法(2D TEE:0.45±0.36 毫米对 2.55±0.99 毫米对 2.59±0.72 毫米,p<0.01;3D TEE:0.34±0.27 毫米对 2.38±0.69 毫米对 1.86±0.77 毫米,p<0.01)。此外,3D TEE 的测量差异比 2D TEE 的内线(2.47±1.86 毫米对 1.86±0.77 毫米,p<0.01)和中线(0.58±0.37 毫米对 0.34±0.27 毫米,p<0.01)方法更准确。

结论

在使用 WATCHMAN FLX 装置进行左心耳封堵时,3D TEE 的中线方法是测量器械最可靠的方法。