Department of Clinical Microbiology, Herlev and Gentofte Hospital , Herlev, Denmark.
Labor Berlin-Charite Vivantes Services GmbH , Berlin, Germany.
J Clin Microbiol. 2023 Oct 24;61(10):e0042623. doi: 10.1128/jcm.00426-23. Epub 2023 Sep 13.
Rapid identification of the causative pathogens of central nervous system infections is essential for providing appropriate management and improving patient outcomes. The performance of QIAstat-Dx Meningitis/Encephalitis (ME) Panel-a multiplex PCR testing platform-in detecting pathogens implicated in meningitis and/or encephalitis was evaluated using BioFire FilmArray ME Panel as a comparator method. This multicenter study analyzed 585 retrospective residual cerebrospinal fluid specimens and 367 contrived specimens. The QIAstat-Dx ME Panel showed positive percent agreement (PPA) values of 100% for , , K1, , and / on clinical samples compared to the BioFire FilmArray ME Panel. The PPA values observed for and were 80% and 88.24%, respectively. Negative percent agreement (NPA) values were >99.0% for each of the six bacterial targets and one fungal target tested with clinical samples. One viral target, herpes simplex virus 1, exhibited a PPA value of 100.0%, while the remaining viral targets-human parechovirus, herpes simplex virus 2, human herpes virus 6, and varicella zoster virus-were >90.0%, with the exception of enterovirus, which had a PPA value of 77.8%. The QIAstat-Dx ME Panel detected five true-positive and four true-negative cases compared to BioFire FilmArray ME Panel. The NPA values for all viral pathogens were >99.0%. Overall, the QIAstat-Dx ME Panel showed comparable performance to the BioFire FilmArray ME Panel as a rapid diagnostic tool for community-acquired meningitis and encephalitis.
快速鉴定中枢神经系统感染的病原体对于提供适当的治疗和改善患者预后至关重要。本研究采用对比方法——生物梅里埃 FilmArray 脑膜炎/脑炎(ME)检测试剂盒,评估了多重 PCR 检测平台——QIAGEN QIAstat-Dx ME 检测试剂盒鉴定与脑膜炎和/或脑炎相关病原体的性能。该多中心研究分析了 585 份回顾性脑脊液剩余样本和 367 份模拟样本。与生物梅里埃 FilmArray ME 检测试剂盒相比,QIAGEN QIAstat-Dx ME 检测试剂盒在临床样本中对 、 、K1、 、和 / 的阳性符合率(PPA)值为 100%。对 和 的 PPA 值分别为 80%和 88.24%。对于六种细菌靶标和一种真菌靶标,临床样本的阴性符合率(NPA)值均>99.0%。一个病毒靶标单纯疱疹病毒 1 的 PPA 值为 100.0%,而其余的病毒靶标——人肠道病毒、单纯疱疹病毒 2、人类疱疹病毒 6 和水痘带状疱疹病毒的 PPA 值均>90.0%,除了肠道病毒的 PPA 值为 77.8%。与生物梅里埃 FilmArray ME 检测试剂盒相比,QIAGEN QIAstat-Dx ME 检测试剂盒检测到 5 例真阳性和 4 例真阴性病例。所有病毒病原体的 NPA 值均>99.0%。总的来说,QIAGEN QIAstat-Dx ME 检测试剂盒作为一种快速诊断工具,用于社区获得性脑膜炎和脑炎,其性能与生物梅里埃 FilmArray ME 检测试剂盒相当。