Thermoregulation is important for maintaining homeostasis in the body. It can be easily broken under anesthesia. An appropriate method for measuring core body temperature is needed, especially for elderly patients, because the efficiency of thermoregulation gradually decreases with age. Zero-heat-flux (ZHF) thermometry (SpotOn) is an alternative, noninvasive method for continuous temperature monitoring at the skin surface. The aim of this study was to examine the accuracy and feasibility of using the SpotOn sensor in lower extremity orthopedic surgery in elderly patients aged over 80 years by comparing a SpotOn sensor with 2 other reliable minimally invasive methods: a tympanic membrane thermometer and a bladder thermometer. This study enrolled 45 patients aged over 80 years who were scheduled to undergo lower extremity surgery. Body temperature was measured using a SpotOn sensor, a tympanic membrane thermometer and a bladder thermometer. Agreements between the SpotOn sensor and the other 2 methods were assessed using Bland and Altman plots for repeated measures adjusted for unequal numbers of measurements per patient. Compared with bladder temperature, bias and limits of agreement for SpotOn temperature were 0.07°C ± 0.58°C. Compared with tympanic membrane temperature, bias and limits of agreement for SpotOn temperature were -0.28°C ± 0.61°C. The 3M SpotOn sensor using the ZHF method for patients aged over 80 years undergoing lower extremity surgery showed feasible measurement value and sensitivity.

摘要

体温调节对于维持身体内环境稳定非常重要。在麻醉下,体温调节很容易被打破。因此,需要一种合适的方法来测量核心体温,特别是对于老年患者,因为随着年龄的增长,体温调节的效率会逐渐降低。零热流(ZHF)测温法(SpotOn)是一种替代的、非侵入性的皮肤表面连续温度监测方法。本研究旨在通过比较 SpotOn 传感器与另外两种可靠的微创方法:鼓膜温度计和膀胱温度计,来检验其在 80 岁以上老年下肢骨科手术中的准确性和可行性。本研究纳入了 45 名计划行下肢手术的 80 岁以上患者。使用 SpotOn 传感器、鼓膜温度计和膀胱温度计测量体温。采用 Bland 和 Altman 图对重复测量进行调整,以评估 SpotOn 传感器与其他 2 种方法之间的一致性。与膀胱温度相比,SpotOn 温度的偏差和一致性界限为 0.07°C ± 0.58°C。与鼓膜温度相比,SpotOn 温度的偏差和一致性界限为-0.28°C ± 0.61°C。对于接受下肢手术的 80 岁以上患者,使用 ZHF 法的 3M SpotOn 传感器具有可行的测量值和灵敏度。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/78b0/10589526/888e386975b7/medi-102-e35593-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验