His bundle pacing (HBP) has become popular in recent years as a more physiological alternative to conventional right ventricular pacing. Implantation requires 12‑lead ECG during surgery, which is not readily available in a standard operating room. Often but not always HBP is performed in an electrophysiology lab. EASI is a reduced lead system which enables derived 12‑lead ECG. EASI derived 12‑lead ECGs on modern tablet computers offer a more mobile and lightweight ECG solution which does not obstruct fluoroscopy during implantation. This case series aims to compare standard 12‑lead ECG to EASI derived 12‑lead ECG in patients undergoing HBP implantation.
A total of 11 patients received permanent HBP guided only by fluoroscopy, a pacing system analyzer (Medtronic CareLink SmartSync Device Manager) and EASI derived 12‑lead ECG (CardioSecur Pro). During the first postoperative device interrogation HBP criteria, as defined in the EHRA consensus paper on conduction system pacing, were evaluated with the EASI derived system as well as a standard 12‑lead ECG and compared to each other. There was perfect agreement with regards to these criteria which lead to identical conclusions in all cases.
HBP implantation can be performed with EASI derived 12‑lead ECG instead of conventional 12‑lead ECG. Criteria for discriminating between selective His bundle, non-selective His bundle or myocardial capture alone are clearly visible in the EASI derived ECG leading to the same conclusion when compared to standard 12‑lead ECG. Compared to a conventional 12‑lead ECG the EASI system offers a leaner setup with less visual obstruction on fluoroscopy.
近年来,希氏束起搏(HBP)已成为一种更具生理性的替代传统右心室起搏的方法。植入需要在手术过程中进行 12 导联心电图(ECG),而这在标准手术室中并不容易获得。通常但并非总是在电生理实验室中进行 HBP。EASI 是一种简化导联系统,可实现 12 导联 ECG 的推导。现代平板电脑上的 EASI 衍生 12 导联 ECG 提供了一种更具移动性和轻便性的 ECG 解决方案,在植入过程中不会妨碍荧光透视。本病例系列旨在比较 HBP 植入患者标准 12 导联 ECG 与 EASI 衍生 12 导联 ECG。
共有 11 例患者仅通过荧光透视、起搏系统分析仪(美敦力 CareLink SmartSync 设备管理器)和 EASI 衍生的 12 导联 ECG(CardioSecur Pro)引导进行了永久性 HBP 植入。在首次术后设备询问中,根据 EHRA 关于传导系统起搏的共识文件,使用 EASI 衍生系统以及标准 12 导联 ECG 评估了 HBP 标准,并将其进行了比较。这些标准的结果完全一致,在所有情况下都得出了相同的结论。
可以使用 EASI 衍生的 12 导联 ECG 代替传统的 12 导联 ECG 进行 HBP 植入。在 EASI 衍生 ECG 中,可以清楚地看到区分选择性希氏束、非选择性希氏束或仅心肌捕获的标准,与标准 12 导联 ECG 相比,这会导致相同的结论。与传统的 12 导联 ECG 相比,EASI 系统提供了更精简的设置,在荧光透视下的视觉干扰更少。