The quantitative aspect of photoplethysmography revised.

作者信息

Jespersen L T, Pedersen O L

出版信息

Heart Vessels. 1986;2(3):186-90. doi: 10.1007/BF02128146.

Abstract

The overall transfer function of a photoplethysmograph (PPG) converting changes in intravascular volume to changes in recorder pen deflection was investigated. The specifications of the apparatus working as a red cell densitometer were assessed in vitro, and it was found that PPG output is a logarithmic function of red cell density roughly in accordance with the Lambert-Beer theorem. In an in vivo experiment it was found that PPG output is also a logarithmic function of digital intravascular volume as assessed by means of air plethysmography. It is suggested that the equation for this in vivo function is different because of a dual action of stasis, producing increases in both intravascular volume and haematocrit. Although the relative contributions of the two disparate phenomena are not assessed further, it is concluded that their joint action on PPG output can be described by a monologarithmic function. In the Appendix, sources of non-linearity in the electronics of PPG are pointed out and instructions for the use of an exponential amplifier which will linearize the overall transfer function are given.

摘要

研究了将血管内容积变化转换为记录仪笔偏转变化的光电容积脉搏波描记仪(PPG)的整体传递函数。对作为红细胞密度计的仪器规格进行了体外评估,发现PPG输出大致符合朗伯-比尔定理,是红细胞密度的对数函数。在一项体内实验中发现,通过空气容积描记法评估,PPG输出也是数字血管内容积的对数函数。有人提出,由于血流淤滞的双重作用,即导致血管内容积和血细胞比容均增加,这种体内功能的方程有所不同。尽管未进一步评估这两种不同现象的相对贡献,但得出结论,它们对PPG输出的联合作用可用单对数函数来描述。在附录中,指出了PPG电子设备中的非线性来源,并给出了使用指数放大器使整体传递函数线性化的说明。