Kunz Sonja, Wang Xiao, Ferrari Uta, Drey Michael, Theodoropoulou Marily, Schilbach Katharina, Reincke Martin, Heier Margit, Peters Annette, Koenig Wolfgang, Zeller Tanja, Thorand Barbara, Bidlingmaier Martin
Department of Medicine IV, University Hospital, Ludwig Maximilian University Munich, Munich, Germany.
Institute of Epidemiology, Helmholtz Zentrum München, German Research Center for Environmental Health, Neuherberg, Germany.
Endocr Connect. 2023 Dec 11;13(1). doi: 10.1530/EC-23-0225. Print 2024 Jan 1.
Measurements utilizing commercially available sets of reagents for determination of steroid hormone profiles by liquid chromatography-tandem mass spectrometry (LC-MS/MS) have become increasingly important for routine laboratories. However, method-specific publications of reference intervals obtained from sufficiently large studies are often missing.
After validation of performance characteristics, a widely available kit for steroid analysis by LC-MS/MS was used to measure concentrations of 15 endogenous steroids (aldosterone, cortisol, cortisone, corticosterone, 11-deoxycortisol, 21-deoxycortisol, dehydroepiandrosterone sulfate, estradiol, testosterone, androstenedione, dihydrotestosterone, dehydroepiandrosterone, 17-hydroxyprogesterone, 11-deoxycorticosterone, progesterone) in more than 500 blood samples from a population-based study. While randomly selected from a larger cohort, the samples equally represented both sexes and covered a wide range of adult age groups. Age- and sex-specific reference intervals were calculated, and correlation with BMI was assessed.
Performance characteristics of the assay matched expectations for 9 of 15 steroids. For most of them, reference intervals obtained from our study population were comparable to those reported by others, with age and sex being the major determinants. A sex-specific correlation with BMI was found for seven steroids. We identified limitations regarding sensitivity of the method for quantification of progesterone in males and postmenopausal females. Concentrations of aldosterone, 21-deoxycortisol, estradiol, 11-deoxycorticosterone, and dihydrotestosterone could not be quantified in a large percentage of samples.
The reference intervals for nine steroids will support meaningful interpretation for steroid profiles as measured by a widely used kit for LC-MS/MS-based quantification. Laboratories using such kits must be aware of potential limitations in sensitivity for some steroids included in the profile.
Quantification of steroid hormones is a cornerstone for diagnosis of several diseases. Commonly used immunoassays have limitations in specificity. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is a promising alternative, particularly if methods are harmonized across laboratories. The use of kits from commercial suppliers might support this. Clinical interpretation of steroid concentrations requires availability of appropriate reference intervals (RIs), but studies on RIs reported in the literature differ in preanalytical and analytical procedures. Here, we provide RIs for steroids measured by a widely available kit under preanalytical conditions mirroring common clinical practice. Such RIs might facilitate interpretation for those using the same method and comparable conditions in clinical routine.
利用市售试剂组通过液相色谱 - 串联质谱法(LC-MS/MS)测定类固醇激素谱,这对常规实验室而言变得愈发重要。然而,通过充分大规模研究获得的特定方法参考区间的出版物常常缺失。
在验证性能特征后,使用一种广泛可用的LC-MS/MS类固醇分析试剂盒,对来自一项基于人群研究的500多个血样中的15种内源性类固醇(醛固酮、皮质醇、可的松、皮质酮、11 - 脱氧皮质醇、21 - 脱氧皮质醇、硫酸脱氢表雄酮、雌二醇、睾酮、雄烯二酮、双氢睾酮、脱氢表雄酮、17 - 羟孕酮、11 - 脱氧皮质酮、孕酮)进行浓度测定。这些样本从一个更大的队列中随机选取,男女比例均衡,涵盖了广泛的成年年龄组。计算了年龄和性别特异性参考区间,并评估了与BMI的相关性。
该检测方法对15种类固醇中的9种的性能特征符合预期。对于其中大多数,我们研究人群获得的参考区间与其他研究报告的区间相当,年龄和性别是主要决定因素。发现7种类固醇与BMI存在性别特异性相关性。我们确定了该方法在男性和绝经后女性中定量孕酮的灵敏度方面存在局限性。在很大比例的样本中,醛固酮、21 - 脱氧皮质醇、雌二醇、11 - 脱氧皮质酮和双氢睾酮的浓度无法定量。
9种类固醇的参考区间将有助于对通过一种广泛使用的基于LC-MS/MS定量的试剂盒所测类固醇谱进行有意义的解读。使用此类试剂盒的实验室必须意识到该谱中某些类固醇在灵敏度方面的潜在局限性。
类固醇激素的定量是多种疾病诊断的基石。常用的免疫测定法在特异性方面存在局限性。液相色谱 - 串联质谱法(LC-MS/MS)是一种有前景的替代方法,特别是如果各实验室的方法能够统一。使用商业供应商的试剂盒可能会有所帮助。类固醇浓度的临床解读需要有合适的参考区间(RI),但文献中关于RI的研究在分析前和分析程序方面存在差异。在此,我们提供了在反映常见临床实践的分析前条件下,通过一种广泛可用的试剂盒所测类固醇的RI。此类RI可能有助于在临床常规中使用相同方法和可比条件的人员进行解读。