J Clin Densitom. 2023 Oct-Dec;26(4):101434. doi: 10.1016/j.jocd.2023.101434. Epub 2023 Oct 28.
The objective of the study was to assess the agreement between the Stratos (DMS) and QDR 4500A (Hologic) DXAs in determining whole body and regional aBMD, as well as whole body composition. Fifty-five individuals (46 women: 84%) with a mean age of 41 ± 13.0 years (range: 20 to 64) and a mean BMI of 31.9 ± 10 kg/m² (range: 12.2 to 49.5) were consecutively scanned on the same day using the two devices. Predictive equations for areal bone mineral density (aBMD) and whole body composition (WBC) were derived from linear regression of the data. The two DXAs were highly correlated (p<0.001 for all parameters) with a correlation coefficient (r) ranging from 0.89 to 0.99 for aBMD (r=0.89 for whole body, r=0.92 for radius, r=0.95 for femoral neck, r=0.96 for total hip, and r=0.99 for L1-L4). For WBC, the r value was 0.98 for lean tissue mass (LTM) and 1.0 for fat mass (FM). Paired t-tests indicated a statistically significant bias between the two DXAs for the majority of measurements, requiring the determination of specific cross-calibration equations. Compared to QDR 4500A, Stratos underestimated whole body aBMD and LTM and overestimated neck and hip aBMD and whole body FM. Conversely, no significant bias was demonstrated for mean aBMD at L1-L4 and radius. For whole body aBMD and FM, the concordance between the two DXAs was influenced by BMI. Despite a high concordance between the two DXAs, the systematic bias for aBMD and WBC measurements illustrates the need to define cross-calibration equations to compare data across systems.
这项研究的目的是评估 Stratos(DMS)和 QDR 4500A(Hologic)DXA 仪在测定全身和局部骨密度(aBMD)以及全身成分方面的一致性。55 名个体(46 名女性:84%),平均年龄 41 ± 13.0 岁(范围:20 至 64 岁),平均 BMI 为 31.9 ± 10kg/m²(范围:12.2 至 49.5kg/m²),在同一天使用两种设备进行连续扫描。通过数据的线性回归得出了用于骨矿物质密度(aBMD)和全身成分(WBC)的预测方程。两种 DXA 仪高度相关(所有参数 p<0.001),aBMD 的相关系数(r)范围为 0.89 至 0.99(全身 r=0.89,桡骨 r=0.92,股骨颈 r=0.95,全髋 r=0.96,L1-L4 r=0.99)。对于 WBC,瘦组织质量(LTM)的 r 值为 0.98,脂肪质量(FM)的 r 值为 1.0。配对 t 检验表明,两种 DXA 仪的大多数测量值均存在统计学上的显著偏差,需要确定特定的交叉校准方程。与 QDR 4500A 相比,Stratos 低估了全身 aBMD 和 LTM,高估了颈部和髋部 aBMD 和全身 FM。相反,L1-L4 和桡骨的平均 aBMD 没有显示出显著的偏差。对于全身 aBMD 和 FM,两种 DXA 仪之间的一致性受到 BMI 的影响。尽管两种 DXA 仪之间具有高度的一致性,但 aBMD 和 WBC 测量的系统偏差表明,需要定义交叉校准方程来比较跨系统的数据。