Successful caudal vena cava and pulmonary vein isolation in healthy horses using 3D electro-anatomical mapping and a contact force-guided ablation system.

机构信息

Equine Cardioteam Ghent, Department of Internal Medicine, Reproduction and Population Medicine, Faculty of Veterinary Medicine, Ghent University, Merelbeke, Belgium.

Department of Cardiology, AZ Sint-Jan, Brugge, Belgium.

出版信息

Equine Vet J. 2024 Sep;56(5):1068-1076. doi: 10.1111/evj.14037. Epub 2023 Dec 27.

Abstract

BACKGROUND

Recently, treatment of equine atrial tachycardia by three-dimensional electro-anatomical mapping (3D EAM) and radiofrequency catheter ablation (RFCA) has been described. Myocardial sleeves in the caudal vena cava and pulmonary veins are a potential trigger for initiation and perpetuation of atrial tachycardia and atrial fibrillation in the horse. Isolation of these myocardial sleeves by RFCA may be an effective treatment for these arrhythmias.

OBJECTIVES

To describe the feasibility of 3D EAM and RFCA to isolate caudal vena cava and pulmonary veins in adult horses using 3D mapping and a contact force (CF)-guided ablation system.

STUDY DESIGN

In vivo experiments.

METHODS

3D EAM and RFCA was performed in five horses without cardiovascular disease under general anaesthesia, using the CF-guided system CARTO®3. Point-by-point RFCA aimed for isolation of caudal vena cava and pulmonary veins. Radiofrequency energy was delivered in power-controlled mode with a target power of 45 W, CF between 10 and 15 g and 30 mL/min irrigation rate, until an ablation-index of 450-500 was reached.

RESULTS

In the right atrium, myocardial sleeves of the caudal vena cava were isolated (n = 5). In the left atrium, isolation of ostium II (n = 3), ostium III (n = 1) and ostium I, II and III en bloc (n = 1) was performed. Successful isolation was confirmed by entrance and exit block.

MAIN LIMITATIONS

Horses were euthanised at the end of the procedure, so long term effects such as potential reconnection of isolated veins could not be studied.

CONCLUSIONS

This is the first description of 3D EAM and RFCA with CARTO®3 in horses, thereby showing the technical feasibility and successful caudal vena cava and pulmonary vein isolation. CF measurement allowed monitoring of catheter-tissue contact, resulting in efficient acute lesion creation as confirmed by entrance and exit block. This is a promising treatment for cardiac arrhythmias in horses.

摘要

背景

最近,使用三维电解剖标测(3D EAM)和射频导管消融(RFCA)治疗马的心房性心动过速已有报道。腔静脉和肺静脉的心肌袖是马心房性心动过速和心房颤动起始和持续的潜在触发因素。通过 RFCA 隔离这些心肌袖可能是治疗这些心律失常的有效方法。

目的

描述使用 3D 标测和接触力(CF)指导消融系统对成年马的腔静脉和肺静脉进行 3D EAM 和 RFCA 隔离的可行性。

研究设计

体内实验。

方法

在全身麻醉下,对 5 匹无心血管疾病的马进行 3D EAM 和 RFCA,使用 CF 指导系统 CARTO®3。通过点到点的 RFCA,旨在隔离腔静脉和肺静脉。以功率控制模式施加射频能量,目标功率为 45 W,CF 为 10-15 g,灌流速度为 30 mL/min,直到达到 450-500 的消融指数。

结果

在右心房,腔静脉的心肌袖被隔离(n=5)。在左心房,成功地隔离了第二(n=3)、第三(n=1)和第一、第二和第三(n=1)的口。通过入口和出口阻滞证实了成功的隔离。

主要局限性

由于在程序结束时对马进行了安乐死,因此无法研究潜在的隔离静脉重新连接等长期影响。

结论

这是首次在马中使用 CARTO®3 进行 3D EAM 和 RFCA 的描述,从而显示了技术的可行性和成功的腔静脉和肺静脉隔离。CF 测量允许监测导管-组织接触,从而有效地创建急性病变,如入口和出口阻滞所证实。这是治疗马心脏心律失常的一种有前途的方法。