Biocompatibility and sub-chronic toxicity studies of phlorotannin/polycaprolactone coated trachea tube for advancing medical device applications.

机构信息

Research Center for Marine-Integrated Bionics Technology, Pukyong National University, Busan, 48513, Republic of Korea.

Marine Integrated Biomedical Technology Center, The National Key Research Institutes in Universities, Pukyong National University, Busan, 48513, Republic of Korea.

出版信息

Sci Rep. 2024 Feb 16;14(1):3945. doi: 10.1038/s41598-024-54684-8.

Abstract

The phlorotannin-polycaprolactone-coated endotracheal tube (PP tube) has been developed with the aim of preventing tracheal stenosis that can result from endotracheal intubation, a factor that can lead to a serious airway obstruction. Its preventive efficacy has been assessed through both in vitro and in vivo investigations. However, there is a lack of studies concerning its biocompatibility and sub-chronic toxicity in animal models, a crucial factor to ensure the safety of its usage as a functional endotracheal tube. Thus, this study aimed to evaluate the biocompatibility and sub-chronic (13 weeks) toxicity of the PP tube through L929 cell line and diverse in vivo models. The cytotoxicity testing was performed using the extracts of PP tube on L929 cells for 72 h. Furthermore, other tests conducted on animal models, including ICR mice (acute systemic toxicity), New Zealand white rabbit (intradermal reactivity and pyrogen tests), guinea pig (maximization sensitization), and Sprague Dawley rats (sub-chronic toxicity). In both biocompatibility and sub-chronic toxicity analyses, no significant adverse effects are observed in the groups exposed to the PP tube, when compared to control group. Altogether, the findings suggested that the PP tube exhibits relative non-toxic and safety, supporting its suitability for clinical usage. However, extended periods of intubation may produce mild irritant responses, highlighting the clinical caution of limiting intubation duration to less than 13 weeks.

摘要

聚己内酯涂层的 phlorotannin 气管导管(PP 管)的开发旨在预防因气管插管引起的气管狭窄,这可能导致严重的气道阻塞。其预防效果已通过体外和体内研究进行了评估。然而,目前缺乏关于其在动物模型中的生物相容性和亚慢性毒性的研究,这是确保其作为功能性气管导管使用的安全性的关键因素。因此,本研究旨在通过 L929 细胞系和多种体内模型评估 PP 管的生物相容性和亚慢性(13 周)毒性。通过将 PP 管提取物作用于 L929 细胞 72 小时来进行细胞毒性测试。此外,还在动物模型上进行了其他测试,包括 ICR 小鼠(急性全身毒性)、新西兰白兔(皮内反应和热原试验)、豚鼠(最大致敏)和 Sprague Dawley 大鼠(亚慢性毒性)。在生物相容性和亚慢性毒性分析中,与对照组相比,暴露于 PP 管的各组均未观察到明显的不良影响。总的来说,这些发现表明 PP 管表现出相对无毒和安全性,支持其在临床应用中的适用性。然而,长时间插管可能会产生轻度刺激性反应,这突显了临床谨慎性,即限制插管时间不超过 13 周。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/f5d0/10873353/906b10daa278/41598_2024_54684_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验