Noda Y, Takai Y, Suto T, Yamada N, Mori T, Kawai N, Kaga T, Hyodo F, Kato H, Matsuo M
Department of Radiology, Gifu University, 1-1 Yanagido, Gifu 501-1194, Japan.
Department of Radiology, Gifu University, 1-1 Yanagido, Gifu 501-1194, Japan.
Clin Radiol. 2024 Apr;79(4):e554-e559. doi: 10.1016/j.crad.2023.12.024. Epub 2024 Jan 18.
To compare the radiation dose, image quality, and conspicuity of pancreatic ductal adenocarcinoma (PDAC) in pancreatic protocol dual-energy computed tomography (CT) between two X-ray tubes mounted in the same CT machine.
This retrospective study comprised 80 patients (median age, 73 years; 45 men) who underwent pancreatic protocol dual-energy CT from January 2019 to March 2022 using either old (Group A, n=41) or new (Group B, n=39) X-ray tubes mounted in the same CT machine. The imaging parameters were completely matched between the two groups, and CT data were reconstructed at 70 and 40 keV. The CT dose-index volume (CTDI); CT attenuation of the abdominal aorta, pancreas, and PDAC; background noise; and qualitative scores for the image noise, overall image quality, and PDAC conspicuity were compared between the two groups.
The CTDI was lower in Group B than Group A (7.9 versus 9.2 mGy; p<0.001). The CT attenuation of all anatomical structures at 70 and 40 keV was comparable between the two groups (p=0.06-0.78). The background noise was lower in Group B than Group A (12 versus 14 HU at 70 keV, p=0.046; and 26 versus 30 HU at 40 keV, p<0.001). Qualitative scores for image noise and overall image quality at 70 and 40 keV and PDAC conspicuity at 40 keV were higher in Group B than Group A (p<0.001-0.045).
The latest X-ray tube could reduce the radiation dose and improve image quality in pancreatic protocol dual-energy CT.
比较安装在同一台CT机上的两个X射线管在胰腺双能量计算机断层扫描(CT)中对胰腺导管腺癌(PDAC)的辐射剂量、图像质量和可探测性。
这项回顾性研究纳入了80例患者(中位年龄73岁;45名男性),他们在2019年1月至2022年3月期间使用安装在同一台CT机上的旧X射线管(A组,n = 41)或新X射线管(B组,n = 39)进行了胰腺双能量CT检查。两组的成像参数完全匹配,并在70 keV和40 keV下重建CT数据。比较两组的CT剂量指数体积(CTDI)、腹主动脉、胰腺和PDAC的CT衰减、背景噪声以及图像噪声、整体图像质量和PDAC可探测性的定性评分。
B组的CTDI低于A组(7.9 mGy对9.2 mGy;p < 0.001)。两组在70 keV和40 keV时所有解剖结构的CT衰减相当(p = 0.06 - 0.78)。B组的背景噪声低于A组(70 keV时为12 HU对14 HU,p = 0.046;40 keV时为26 HU对30 HU,p < 0.001)。B组在70 keV和40 keV时的图像噪声和整体图像质量的定性评分以及40 keV时的PDAC可探测性评分高于A组(p < 0.001 - 0.045)。
最新的X射线管可降低胰腺双能量CT的辐射剂量并提高图像质量。