J Dent. 2024 May;144:104943. doi: 10.1016/j.jdent.2024.104943. Epub 2024 Mar 15.
This study aimed to evaluate the accuracy of an intraoral scanner (IOS - Medit i700) on tooth abutments with vertical preparations at 2 depths below the free gingival margin, and to determine if the IOS can reproduce the area beyond the finish surface of the tested preparation geometry.
Two abutments for a maxillary first molar were designed by means of CAD software, with vertical preparations set at 1 and 2 mm below the gingiva. These abutments were subsequently printed in resin and placed on a reference model. The reference files consisted of scans made using a metrological machine on these abutments. Ten scans were made with the tested IOS on each sample, resulting in two study groups. The scans from the experimental groups were labeled "V-1″ for vertical preparation at 1 mm below the gingival margin and "V-2″ for 2 mm below. The analysis of these scans was performed using Geomagic Control X (3D SYSTEMS) to assess their trueness and precision in µm. Descriptive statistics with a 95 % confidence interval were employed, alongside independent sample tests, to ascertain any differences between the groups (α=0.05).
Statistically significant differences were not found both for trueness (p=.104) and precision (p=.409), between the tested geometries. The mean values for trueness were V-1 = 37.5[31.4-43.6]; V-2 = 32.6[30.6-34.6]. About the precision, the mean values were V-1 = 20.5[8.4-32.5]; V-2 = 18.4[8.2-28.5]. In both the study groups, it was possible to detect the surface beyond the finish area.
Within the limitations of this study, vertical preparation design allows for registration of the tooth anatomy beyond the finish area with IOS. Moreover, the mean accuracy values were clinically acceptable at both 1 and 2 mm below the gingival margin.
本研究旨在评估口内扫描仪(IOS - Medit i700)在牙龈缘下 2 个深度的垂直预备体上的准确性,并确定该扫描仪是否能再现测试预备体几何形状的完成面以外的区域。
通过 CAD 软件设计上颌第一磨牙的两个基牙,预备体垂直深度分别设置在牙龈下 1 和 2 毫米处。然后,这些基牙被打印在树脂中并放置在参考模型上。参考文件由在这些基牙上使用计量机进行的扫描组成。每个样本用测试 IOS 进行了 10 次扫描,共分为两组。实验组的扫描标记为牙龈缘下 1 毫米处的垂直预备体为“V-1”,标记为牙龈缘下 2 毫米处的垂直预备体为“V-2”。使用 Geomagic Control X(3D SYSTEMS)对这些扫描进行分析,以评估其在 µm 下的准确性和精度。采用描述性统计和 95%置信区间,以及独立样本检验,以确定组间的任何差异(α=0.05)。
在所测试的几何形状中,无论是准确性(p=.104)还是精度(p=.409),两组之间均未发现统计学差异。准确性的平均值为 V-1=37.5[31.4-43.6];V-2=32.6[30.6-34.6]。关于精度,平均值为 V-1=20.5[8.4-32.5];V-2=18.4[8.2-28.5]。在两个研究组中,都可以检测到完成区域以外的表面。
在本研究的限制范围内,垂直预备设计允许 IOS 记录完成区域以外的牙体解剖结构。此外,在牙龈缘下 1 毫米和 2 毫米处,平均准确性值均在临床可接受范围内。