无导线起搏器植入后反复出现与频率相关的起搏阈值升高 1 例。

A case of repeated rate-dependent pacing threshold elevation after leadless pacemaker implantation.

机构信息

Department of Cardiovascular Medicine, Osaka Metropolitan University Graduate School of Medicine, Osaka, Japan.

出版信息

Pacing Clin Electrophysiol. 2024 Oct;47(10):1391-1393. doi: 10.1111/pace.14999. Epub 2024 May 18.

Abstract

The pacing threshold is important for leadless pacemakers, as the pacing output has a significant impact on battery longevity. Acute pacing rate-dependent threshold increases have also been reported with leadless pacemakers. In the present case, we experienced a case in which the threshold, which had been raised in the acute phase, once showed a tendency to improve, but then worsened again. And, as in previous reports, thresholds improved completely in the chronic phase. Repeated retrieval and reimplantation of leadless pacemakers increases the likelihood of fatal complications, so being aware of such phenomenon may prevent unnecessary procedure and complications.

摘要

起搏阈值对于无导线起搏器很重要,因为起搏输出对电池寿命有重大影响。无导线起搏器也有报道急性起搏频率依赖性阈值升高的情况。在本病例中,我们遇到了一个在急性期升高的阈值曾经有改善的趋势,但后来又再次恶化的病例。并且,与之前的报告一样,阈值在慢性期完全改善。无导线起搏器的反复取出和重新植入增加了致命并发症的可能性,因此了解这种现象可能会预防不必要的程序和并发症。