在Halcyon设备上对孤立性肺病变单次立体定向给予30 Gy剂量的可行性。

Feasibility of a Single-Fraction Stereotactic Dose of 30 Gy to Solitary Lung Lesions on Halcyon.

作者信息

Misa Joshua, Knight James A, Pokhrel Damodar

机构信息

Department of Radiation Oncology, University of Kentucky, Lexington, USA.

出版信息

Cureus. 2024 May 2;16(5):e59535. doi: 10.7759/cureus.59535. eCollection 2024 May.

Abstract

Purpose We sought to explore the feasibility of using the current co-planar Halcyon ring delivery system (RDS) with a novel multileaf collimator (MLC) aperture shape controller in delivering a single high dose of 30 Gy to solitary lung lesions via stereotactic body radiotherapy (SBRT). Materials and methods Thirteen non-small-cell lung cancer (NSCLC) patients previously treated with a single dose of 30 Gy to lung lesions via SBRT on the TrueBeam (6MV-FFF) using non-coplanar volumetric modulated arc therapy (VMAT) arcs were anonymized and replanned onto the Halcyon RDS (6MV-FFF) following RTOG-0915 single-fraction criteria. The Halcyon plans utilized a novel dynamic conformal arc (DCA)-based MLC-fitting approach before VMAT optimization with a user-defined aperture shape controller option. The clinical TrueBeam and Halcyon plans were compared via their protocol compliance, target conformity, gradient index, and dose to organs-at-risk (OAR). Treatment delivery efficacy and accuracy were assessed through end-to-end quality assurance (QA) tests on Halcyon and independent dose verification via in-house Monte Carlo (MC) second-check validation. Results All Halcyon lung SBRT plans met RTOG-0915 protocol's requirements for target coverage, conformity, and gradient indices, and maximum dose 2 cm away from the target (D) while being statistically insignificant (p > 0.05) when compared to clinical TrueBeam plans. Additionally, Halcyon provided a similar dose to OAR except for the ribs, where Halcyon demonstrated a lower maximum dose (15.22 Gy vs 17.01 Gy, p < 0.001). However, Halcyon plans required a higher total monitor unit (8892 MU vs 7413 MU, p < 0.001), resulting in a higher beam modulation factor (2.96 MU/cGy vs 2.47 MU/cGy, p < 0.001) and an increase in beam-on time by a factor of 2.1 (11.11 min vs 5.3 min, p < 0.005). End-to-end QA measurements demonstrate that Halcyon plans were clinically acceptable with an average gamma passing rate of 99.8% for 2%/2mm criteria and independent MC 2nd checks within ±2.86%. Conclusion Our end-to-end testing and validation study demonstrates that by utilizing a DCA-based MLC aperture shape controller before VMAT optimization, Halcyon can be used for delivering a single dose of lung SBRT treatment. However, future improvements of Halcyon RDS are recommended to allow higher output rates, rotational couch corrections, and an integrated intrafraction motion management system that will further enhance Halcyon's capability for site-specific single dosage of SBRT.

摘要

目的 我们试图探讨使用当前的共面Halcyon环形放射治疗系统(RDS)与一种新型多叶准直器(MLC)孔径形状控制器,通过立体定向体部放射治疗(SBRT)向孤立性肺病变单次给予30 Gy高剂量照射的可行性。材料与方法 对13例非小细胞肺癌(NSCLC)患者进行匿名处理,这些患者之前在TrueBeam(6MV-FFF)上采用非共面容积调强弧形治疗(VMAT)弧对肺部病变进行了单次30 Gy的SBRT治疗,并按照RTOG-0915单分割标准在Halcyon RDS(6MV-FFF)上重新进行计划。Halcyon计划在VMAT优化之前采用一种基于新型动态适形弧(DCA)的MLC拟合方法,并使用用户定义的孔径形状控制器选项。通过方案依从性、靶区适形性、梯度指数以及危及器官(OAR)的剂量,对临床TrueBeam计划和Halcyon计划进行比较。通过对Halcyon进行端到端质量保证(QA)测试以及通过内部蒙特卡罗(MC)二次检查验证进行独立剂量验证,评估治疗实施的有效性和准确性。结果 所有Halcyon肺部SBRT计划均符合RTOG-0915方案对靶区覆盖、适形性和梯度指数的要求,以及距靶区2 cm处的最大剂量要求(D),与临床TrueBeam计划相比,差异无统计学意义(p>0.05)。此外,除肋骨外,Halcyon对OAR的剂量相似,在肋骨处Halcyon的最大剂量较低(15.22 Gy对17.01 Gy,p<0.001)。然而,Halcyon计划需要更高的总监测单位(8892 MU对7413 MU,p<0.001),导致更高的射束调制因子(2.96 MU/cGy对2.47 MU/cGy,p<0.001),并且射束开启时间增加了2.1倍(11.11分钟对5.3分钟,p<0.005)。端到端QA测量表明,Halcyon计划在临床上是可接受的,对于2%/2mm标准,平均伽马通过率为99.8%,独立MC二次检查在±2.86%以内。结论 我们的端到端测试和验证研究表明,通过在VMAT优化之前使用基于DCA的MLC孔径形状控制器,Halcyon可用于实施单次肺部SBRT治疗。然而,建议对Halcyon RDS进行未来改进,以实现更高的输出率、旋转治疗床校正以及集成的分次内运动管理系统,这将进一步提高Halcyon针对特定部位单次剂量SBRT的能力。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/84f6/11144037/7f0f6995f3a9/cureus-0016-00000059535-i01.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验