Kim J H, Lee B S, Park S B
Ultrason Imaging. 1985 Apr;7(2):179-90. doi: 10.1177/016173468500700207.
A prior inverse filtering technique is applied to improve the axial resolution in ultrasonic imaging. The ultrasound transducer is excited by a synthesized signal corresponding to the impulse response of the inverse filter. The diffraction effects of the transducer and the observed scatterer, which limit the performance of the inverse filter, are examined experimentally as well as theoretically. Using a B-scan image of a tissue-equivalent phantom, we demonstrate the possibility of applying the prior inverse filtering method to improve medical ultrasound images.
一种先验逆滤波技术被应用于提高超声成像中的轴向分辨率。超声换能器由对应于逆滤波器脉冲响应的合成信号激励。换能器和观测散射体的衍射效应限制了逆滤波器的性能,本文从实验和理论两方面对其进行了研究。利用组织等效体模的B扫描图像,我们证明了应用先验逆滤波方法改善医学超声图像的可能性。