Precise Determination of Cd Isotope Ratios at 3-10 ng Level by Thermal Ionization Mass Spectrometry Using a Molybdenum Silicide Emitter.

作者信息

Li Chao-Feng, Chu Zhu-Yin, Peng Peng

机构信息

State Key Laboratory of Lithospheric Evolution, Institute of Geology and Geophysics, Chinese Academy of Sciences, Beijing 100029, China.

Innovation Academy for Earth Science, Chinese Academy of Sciences, Beijing 100029, China.

出版信息

Anal Chem. 2024 Sep 10. doi: 10.1021/acs.analchem.4c03362.

Abstract

Thermal ionization mass spectrometry (TIMS) combined with the double spike technique has excellent analytical precision for Cd isotopic ratio analysis. However, because of the low ionization efficiency of Cd by TIMS, it is still not possible to obtain high precision Cd isotope ratios for small sample size (<100 ng) due to the lack of a highly sensitive emitter for Cd. A new loading method using molybdenum silicide (MoSi) emitter has been developed for Cd isotope ratio measurements. This emitter produces a significant enhancement in the ionization efficiency of Cd and thus significantly reduces the required sample size to the 3-10 ng level. Analyses of δCd for the NIST SRM 3108 using Cd-Cd double spike method show excellent reproducibility (2 SD) that reaches ±0.032‰, ±0.042‰, and ±0.051‰ for 10, 5, and 3 ng of Cd, respectively. This method was further verified with a suite of geological reference materials. Replicate digestions and analyses ( = 8, 2 SD) of δCd for NIST SRM 2711a, NOD A-1, and GBW08401 demonstrated good external reproducibility with results of 0.596 ± 0.024‰ for NIST SRM 2711a, 0.150 ± 0.036‰ for NOD A-1, and -0.665 ± 0.084‰ for GBW08401. These data clearly indicate that MoSi is an excellent alternative for traditional silica gel to Cd isotopic measurements, especially for samples with a low content of Cd.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验