Panichapat Nuttida, Niyomkarn Watit, Boonjindasup Wicharn, Thiamrakij Pariyapa, Sritippayawan Suchada, Deerojanawong Jitladda
Division of Pulmonology, Department of Pediatrics, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Division of Neurology, Department of Pediatrics, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
J Clin Sleep Med. 2025 Jan 1;21(1):123-128. doi: 10.5664/jcsm.11348.
The Belun ring is a new home sleep apnea testing device using a pulse oximeter sensor and a neural network algorithm, but its data in children are limited. This study aims to evaluate the correlation and agreement of the Belun ring, compared with polysomnography (PSG) and determine the diagnostic accuracy of the Belun ring for moderate-to-severe obstructive sleep apnea (OSA).
This is a cross-sectional observational study in children aged 5-18 years with suspected OSA between June 2023 and February 2024. The Belun ring and PSG were undertaken on eligible participants to assess apnea-hypopnea index (AHI) in the same sleep test session.
Of 75 children enrolled, OSA was diagnosed in 74 children by PSG. The Belun AHI (B-AHI) was moderately correlated with the PSG AHI (P-AHI) ( .63, < .001) with mean difference (standard deviation) -7.8 (13.91) events/h. The area under the receiver operating characteristic curve of the B-AHI to identify moderate-to-severe OSA (P-AHI > 5 events/h) was 0.66, and the B-AHI cut-off of 3 events/h yielded 74.1% sensitivity and 52.4% specificity. The B-AHI cut-off of 2 events/h yielded 92.6% sensitivity, and 7 events/h yielded 95.2% specificity.
Despite the correlation, the difference in AHI between the Belun ring and PSG in children was noted. Either single or multiple B-AHI cut-offs to diagnose, include or exclude moderate-to-severe OSA may be valuable, but their implementation must be approached with caution.
Registry: Thai Clinical Trials Registry; Name: Diagnostic Accuracy of the Belun Ring in Children at Risk of Obstructive Sleep Apnea; URL: https://www.thaiclinicaltrials.org/show/TCTR20240604003; Identifier: TCTR20240604003.
Panichapat N, Niyomkarn W, Boonjindasup W, Thiamrakij P, Sritippayawan S, Deerojanawong J. Diagnostic accuracy of the Belun ring in children at risk of obstructive sleep apnea. . 2025;21(1):123-128.
贝伦环是一种使用脉搏血氧仪传感器和神经网络算法的新型家庭睡眠呼吸暂停检测设备,但关于其在儿童中的数据有限。本研究旨在评估贝伦环与多导睡眠图(PSG)的相关性和一致性,并确定贝伦环对中重度阻塞性睡眠呼吸暂停(OSA)的诊断准确性。
这是一项横断面观察性研究,研究对象为2023年6月至2024年2月期间年龄在5至18岁、疑似患有OSA的儿童。对符合条件的参与者进行贝伦环和PSG检测,以在同一睡眠测试时段评估呼吸暂停低通气指数(AHI)。
在纳入的75名儿童中,PSG诊断出74名儿童患有OSA。贝伦AHI(B-AHI)与PSG AHI(P-AHI)呈中度相关(r = 0.63,P < 0.001),平均差异(标准差)为-7.8(13.91)次/小时。B-AHI识别中重度OSA(P-AHI > 5次/小时)的受试者工作特征曲线下面积为0.66,B-AHI截断值为3次/小时时,敏感性为74.1%,特异性为52.4%。B-AHI截断值为2次/小时时,敏感性为92.6%,截断值为7次/小时时,特异性为95.2%。
尽管存在相关性,但注意到儿童中贝伦环和PSG的AHI存在差异。用于诊断、纳入或排除中重度OSA的单一或多个B-AHI截断值可能有价值,但在实施时必须谨慎。
注册机构:泰国临床试验注册中心;名称:贝伦环对有阻塞性睡眠呼吸暂停风险儿童的诊断准确性;网址:https://www.thaiclinicaltrials.org/show/TCTR20240604003;标识符:TCTR20240604003。
Panichapat N, Niyomkarn W, Boonjindasup W, Thiamrakij P, Sritippayawan S, Deerojanawong J. 贝伦环对有阻塞性睡眠呼吸暂停风险儿童的诊断准确性。……2025;21(1):123 - 128。