Schmitt Katharina, Timm Malena, Krüger Philipp, Oppel Niels, Napp Alexandra, Pohl Friederike, Schuon Robert, Kötter Lisa, Bankstahl Marion, Lenarz Thomas, Stein Tobias, Paasche Gerrit
Department of Otorhinolaryngology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany.
bess pro GmbH, Gustav-Krone-Str. 7, 14167 Berlin, Germany.
Bioengineering (Basel). 2024 Oct 11;11(10):1015. doi: 10.3390/bioengineering11101015.
Unmet needs in the treatment of chronic otitis media and Eustachian tube dysfunction (ETD) triggered the development of stents for the Eustachian tube (ET). In this study, for the first time, stents were placed in an artificially blocked ET to evaluate stent function. Eight adult female sheep were injected with stabilized hyaluronic acid (HA) on both sides to induce ETD. Subsequently, a tapered nitinol ET stent was inserted on one side, and animals were examined bilaterally by endoscopy, tympanometry, cone beam computed tomography, and final histology. Seven of the stents were placed in the desired cartilaginous portion of the ET. At the end of the study, one stented side appeared slightly open; all other ET orifices were closed. Tympanometry revealed re-ventilation of the middle ear in four out of seven correctly stented animals within 3 to 6 weeks after stent insertion. The major amount of HA was found at the pharyngeal orifice of the ET anterior to the stent. Thus, the stent position did not completely align with the HA position. While a functional analysis will require refinement of the experimental setup, this study provides first promising results for stent insertion in a sheep model of ETD.
慢性中耳炎和咽鼓管功能障碍(ETD)治疗中未满足的需求促使了咽鼓管(ET)支架的研发。在本研究中,首次将支架置入人工阻塞的咽鼓管以评估支架功能。对八只成年雌性绵羊双侧注射稳定化透明质酸(HA)以诱发ETD。随后,在一侧插入锥形镍钛合金咽鼓管支架,并通过内窥镜检查、鼓室导抗图、锥形束计算机断层扫描和最终组织学对动物进行双侧检查。七个支架放置在咽鼓管所需的软骨部分。在研究结束时,一个置入支架的一侧略显开放;所有其他咽鼓管口均闭合。鼓室导抗图显示,在七个正确置入支架的动物中,有四个在支架置入后3至6周内中耳重新通气。在支架前方的咽鼓管咽口发现了大量的HA。因此,支架位置与HA位置并未完全对齐。虽然功能分析需要完善实验设置,但本研究为在ETD绵羊模型中置入支架提供了首个有前景的结果。