Shimuta Toru, Hanada Kaname, Ryu Kazuteru, Idei Koichi, Kanai Nobuyuki
Minato MIRAI Innovation Center, Murata Manufacturing Co., Ltd., 4-3-8, Minatomirai, Nishi-ku, Yokohama-shi 220-0012, Japan.
Orthopedic Surgery, Kanai Hospital, 612-12, Yodokizu-cho, Fushimi-ku, Kyoto 613-0911, Japan.
Sensors (Basel). 2024 Dec 10;24(24):7882. doi: 10.3390/s24247882.
Diabetes is known to reduce blood circulation in capillaries and arterioles; however, no devices can easily measure this on a daily basis. In this study, we developed a tool for measuring finger photoplethysmograms using green light and near-infrared LEDs. Thereafter, photoplethysmography was conducted on 25 inpatients/outpatients with diabetes and 21 adult males and females who had not been diagnosed with or treated for diabetes, hypertension, or cardiovascular disease (control group). In patients with diabetes, the inverse full width at half-maximum velocity plethysmogram tended to be smaller than that in the control group, and the delay in the green light a-wave peak relative to the near-infrared light a-wave peak in the acceleration plethysmogram was significantly increased. The results suggest that peripheral hemodynamics can be easily estimated at home using a photoplethysmography device mounted on a ring-wearable device.
众所周知,糖尿病会减少毛细血管和小动脉中的血液循环;然而,目前尚无设备能够轻松地进行日常测量。在本研究中,我们开发了一种使用绿光和近红外发光二极管来测量手指光电容积脉搏波的工具。此后,对25名糖尿病住院患者/门诊患者以及21名未被诊断患有糖尿病、高血压或心血管疾病的成年男性和女性(对照组)进行了光电容积脉搏波描记术。在糖尿病患者中,反向半高宽速度容积脉搏图往往比对照组小,并且在加速度容积脉搏图中,绿光a波峰相对于近红外光a波峰的延迟显著增加。结果表明,使用安装在可穿戴戒指设备上的光电容积脉搏波描记术设备,可以在家中轻松估计外周血流动力学。