种植体扫描体类型、口内扫描仪和扫描策略对上颌全牙弓种植体扫描准确性和扫描时间的影响:一项体外研究。

Effect of Implant Scan Body type, Intraoral scanner and Scan Strategy on the accuracy and scanning time of a maxillary complete arch implant scans: an in vitro study.

作者信息

Grande Francesco, Mosca Balma Alessandro, Mussano Federico, Catapano Santo

机构信息

PhD student, Department of Mechanical and aerospace engineering, Polytechnic University of Turin, Turin, Italy.

Biomedical Engineer, Department of Surgical Sciences, CIR Dental School, University of Turin, Turin, Italy.

出版信息

J Dent. 2025 Aug;159:105782. doi: 10.1016/j.jdent.2025.105782. Epub 2025 Apr 25.

Abstract

OBJECTIVES

This study aimed to investigate the influence of three Implant Scan Bodies (ISBs), four intraoral scanners (IOSs), and two scan strategies on the trueness of complete arch implant scans.

METHODS

A reference digital cast of an all-on-six maxillary rehabilitation was obtained using a coordinate measuring machine. Three ISBs - AQ (ISB-AQ), One-shot, Elos (ISB-PEEK) - four IOSs - Trios3POD (TRIOS), Medit i700 (MED), iTero Element 5D (iTe), Primescan (PRIME) - and two scan strategies - Zig-zag (ZZ), One-shot (OS) - were used for evaluation. For each group created by the combination of the different variables, 10 digital scans were taken by one operator (n = 240) under controlled temperature and lighting conditions. The positions of the ISB connections were compared to the standard tessellation language reference file by using the nurbs-to-nurbs method. The overall 3D linear and angular deviations of the ISBs connections between reference and test scans were evaluated by computing the Euler distances (α=0.05). Interimplant distance discrepancies were calculated at the ISB connection level. Scanning times were recorded.

RESULTS

Regarding linear and angular deviations, AQ had significantly higher trueness than the other ISBs (p < 0.01). TRIOS showed significantly higher trueness (p < 0.01). No significant differences for scan strategies were found (p > 0.05). ISB-OneSh, Primescan and ZZ method resulted in significant shortest scanning times (p<<0.01).

CONCLUSIONS

ISB-AQ demonstrated higher trueness than One-Shot and PEEK. TRIOS performed better than the other IOSs. Scan strategy didn't influence the trueness. ISB-OneSh, Primescan, and ZZ strategy allowed the fastest scans.

CLINICAL SIGNIFICANCE

The choice of ISB and IOS significantly affects the accuracy of complete-arch implant scans. The scan strategy did not impact trueness but influenced the efficiency. As an in vitro study, results observed need confirmation in the clinical settings.

摘要

目的

本研究旨在探讨三种种植体扫描体(ISB)、四种口内扫描仪(IOS)和两种扫描策略对全牙弓种植体扫描准确性的影响。

方法

使用坐标测量机获取全口六颗上颌种植修复体的参考数字模型。使用三种ISB——AQ(ISB-AQ)、一次性扫描体、Elos(ISB-PEEK),四种IOS——Trios3POD(TRIOS)、Medit i700(MED)、iTero Element 5D(iTe)、Primescan(PRIME),以及两种扫描策略——之字形(ZZ)、一次性扫描(OS)进行评估。对于由不同变量组合形成的每组,由一名操作人员在受控的温度和光照条件下进行10次数字扫描(n = 240)。通过使用非均匀有理B样条曲线到非均匀有理B样条曲线的方法,将ISB连接的位置与标准镶嵌语言参考文件进行比较。通过计算欧拉距离评估参考扫描和测试扫描之间ISB连接的总体三维线性和角度偏差(α = 0.05)。在ISB连接水平计算种植体间距离差异。记录扫描时间。

结果

关于线性和角度偏差,AQ的准确性显著高于其他ISB(p < 0.01)。TRIOS显示出显著更高的准确性(p < 0.01)。未发现扫描策略存在显著差异(p > 0.05)。ISB-OneSh、Primescan和ZZ方法的扫描时间显著最短(p << 0.01)。

结论

ISB-AQ的准确性高于One-Shot和PEEK。TRIOS的表现优于其他IOS。扫描策略不影响准确性。ISB-OneSh、Primescan和ZZ策略可实现最快扫描。

临床意义

ISB和IOS的选择显著影响全牙弓种植体扫描的准确性。扫描策略不影响准确性,但影响效率。作为一项体外研究,观察到的结果需要在临床环境中得到证实。