BACKGROUND

Breast implant-associated anaplastic large cell lymphoma (BIA-ALCL) has affected more than 1700 women. Treatment morbidity is reduced and survival improved by early detection and removal of implant and capsule.

OBJECTIVE

To address the need for early detection, we developed a multiplex lateral flow assay (LFA) for soluble IL-10 and CD30 in seromas surrounding implants.

METHODS

The range of detection was determined with recombinant CD30 or IL-10 added to a benign seroma to simulate clinical conditions. Then clinical samples from 23 women with BIA-ALCL and 17 benign seromas were analyzed with a multiplex LFA detecting soluble IL-10 and CD30.

RESULTS

When recombinant proteins were added to benign seromas, CD30 was detected down to a concentration of 0.5 ng/mL, and IL-10 could be detected to down to 5 ng/mL. Significant differences were found between CD30 and IL-10 test line/control line ratios (TL/CL) of 23 BIA-ALCL and 17 benign seromas. Visual inspection of LFAs resulted in 65% sensitivity, and improved to 87% with image analysis.

CONCLUSIONS

We conclude that the multiplex IL-10/CD30 can detect BIA-ALCL with 100% specificity and 65% sensitivity in this group of 40 clinical seromas based on visual assessment of LFA strips. Sensitivity was improved to 87% with image analysis, which can provide a screening test for BIA-ALCL. The multiplex LFA demonstrated a perfect positive predictive value (100%), indicating that all positive results correctly identified BIA-ALCL cases. The negative predictive value of 99% can exclude BIA-ALCL.

摘要

背景

乳房植入物相关间变性大细胞淋巴瘤(BIA-ALCL)已影响了1700多名女性。通过早期发现并移除植入物和包膜,可降低治疗的发病率并提高生存率。

目的

为满足早期检测的需求,我们开发了一种用于检测植入物周围血清肿中可溶性白细胞介素-10(IL-10)和CD30的多重侧向流动分析(LFA)方法。

方法

将重组CD30或IL-10添加到良性血清肿中以模拟临床情况,从而确定检测范围。然后,使用检测可溶性IL-10和CD30的多重LFA对23例BIA-ALCL女性患者的临床样本和17例良性血清肿进行分析。

结果

当将重组蛋白添加到良性血清肿中时,可检测到低至0.5 ng/mL浓度的CD30,以及低至5 ng/mL浓度的IL-10。在23例BIA-ALCL患者和17例良性血清肿的CD30和IL-10检测线/对照线比率(TL/CL)之间发现了显著差异。对LFA进行目视检查的灵敏度为65%,通过图像分析可提高到87%。

结论

我们得出结论,基于对LFA试纸条的目视评估,在这40例临床血清肿中,多重IL-10/CD30检测BIA-ALCL的特异性为100%,灵敏度为65%。通过图像分析,灵敏度提高到87%,可为BIA-ALCL提供一种筛查检测方法。多重LFA显示出完美的阳性预测值(100%),表明所有阳性结果均能正确识别BIA-ALCL病例。99%的阴性预测值可排除BIA-ALCL。