Ashton Christopher B J, Hammer Arthur, Heeren Tjebo F C, Allan Bruce D, Maurino Vincenzo, Gore Daniel M
J Refract Surg. 2025 Jul;41(7):e702-e708. doi: 10.3928/1081597X-20250520-02. Epub 2025 Jul 1.
To compare clinically achieved vaults with the predicted NK-formula V2 (NK-V2) and NK-formula V3 (NK-V3) vaults in a White and Asian population in patients undergoing Implantable Collamer Lens (ICL) surgery.
Data were collected retrospectively from three surgeons for all patients who had ICL surgery for myopia between September 2020 and August 2022 at Moorfields Eye Hospital. ICL sizing was performed using the optimal NK-V2 sizing formula. The postoperative vault was measured using anterior segment optical coherence tomography. Back calculation was used to evaluate predicted versus achieved vaults for the newer NK-V3 formula.
A total of 136 eyes from 72 patients were eligible. Seventy-two eyes of 72 patients were included in the analysis, 50 White (69%) and 22 Asian (31%). The mean residual vault for NK-V2 was -0.08 and -0.11 ( = .76) and NK-V3 was 0.04 and 0.02 ( = .78) for White and Asian, respectively. Mean absolute errors of the White sample were 0.2 and 0.16 for NKV2 and NK-V3, respectively ( = .23). Mean absolute errors of the Asian sample were 0.31 and 0.22 for NK-V2 and NK-V3, respectively ( = .16). There was no statistical significance when comparing mean absolute errors from White and Asian samples for NK-V2 or NK-V3.
NK-V2 formula for ICL sizing, originally derived from an Asian population sample, appeared to perform at least as well in Whites with no adverse outcomes or safety issues. NK-V3 was superior in predicting postoperative vault compared to NK-V2 in Asian eyes and comparable for White eyes. .
在接受可植入式胶原晶状体(ICL)手术的白种人和亚洲人群中,比较临床获得的房水腔与预测的NK公式V2(NK-V2)和NK公式V3(NK-V3)房水腔。
回顾性收集2020年9月至2022年8月在摩尔菲尔德眼科医院接受近视ICL手术的所有患者的数据,这些数据来自三位外科医生。使用最佳NK-V2尺寸计算公式进行ICL尺寸测量。使用眼前节光学相干断层扫描测量术后房水腔。采用反向计算来评估新的NK-V3公式预测的房水腔与实际获得的房水腔。
共有72例患者的136只眼符合条件。72例患者的72只眼纳入分析,其中50只为白种人(69%),22只为亚洲人(31%)。白种人和亚洲人的NK-V2平均残余房水腔分别为-0.08和-0.11(P = 0.76),NK-V3分别为0.04和0.02(P = 0.78)。白种人样本中,NK-V2和NK-V3的平均绝对误差分别为0.2和0.16(P = 0.23)。亚洲人样本中,NK-V2和NK-V3的平均绝对误差分别为0.31和0.22(P = 0.16)。比较白种人和亚洲人样本中NK-V2或NK-V3的平均绝对误差时,无统计学意义。
最初源自亚洲人群样本的ICL尺寸计算NK-V2公式,在白种人中表现至少同样良好,且无不良后果或安全问题。在亚洲眼中,NK-V3在预测术后房水腔方面优于NK-V2,在白种眼中两者相当。