Abstract

OBJECTIVES

This in vitro study aimed to evaluate temperature variation on the external surface of mandibular molars and within ultrasonic inserts when using adapter-coupled versus one-piece inserts.

METHODS

Twenty-four extracted human mandibular molars were divided into two groups based on the type of ultrasonic insert used: adapter-coupled and one-piece inserts. Temperature on the external surface of each tooth was measured with a thermocouple probe positioned in the furcation area, capturing data continuously. The temperature of the ultrasonic inserts was monitored in real-time using a thermal imaging camera. Measurements were taken in a controlled environment without cooling for over 120 seconds. Statistical analysis was conducted using analysis of variance (ANOVA) and two-way ANOVA with repeated measures to evaluate temperature variations between groups and over time, with significance set at 5%.

RESULTS

In the external tooth surface temperature measurements, no significant differences were observed between the groups during the initial 15 seconds (p = 0.185) and 30 seconds (p = 0.067). However, significant differences emerged at 60 seconds (p = 0.025), 90 seconds (p = 0.024), and 120 seconds (p = 0.020), with the one-piece insert group demonstrating higher temperatures in the furcation region. Thermal imaging of the inserts revealed a significant difference at all time points (p < 0.001), with adapter-coupled inserts showing greater heating.

CONCLUSIONS

The use of ultrasonic inserts leads to a gradual rise in temperature on the external tooth surface. One-piece inserts generated higher temperatures on the tooth, while adapter-coupled inserts exhibited greater heating within the insert.

摘要

目的

本体外研究旨在评估使用转接耦合式与一体式超声插入器械时,下颌磨牙外表面及超声插入器械内部的温度变化。

方法

根据所使用的超声插入器械类型,将24颗拔除的人下颌磨牙分为两组:转接耦合式和一体式插入器械。使用置于根分叉区的热电偶探头测量每颗牙齿外表面的温度,持续采集数据。使用热成像相机实时监测超声插入器械的温度。在无冷却的受控环境中进行测量,持续时间超过120秒。采用方差分析(ANOVA)和重复测量的双向ANOVA进行统计分析,以评估组间和随时间的温度变化,显著性设定为5%。

结果

在牙齿外表面温度测量中,两组在最初15秒(p = 0.185)和30秒(p = 0.067)时未观察到显著差异。然而,在60秒(p = 0.025)、90秒(p = 0.024)和120秒(p = 0.020)时出现了显著差异,一体式插入器械组在根分叉区域显示出更高的温度。插入器械的热成像显示在所有时间点均存在显著差异(p < 0.001),转接耦合式插入器械显示出更大的发热。

结论

使用超声插入器械会导致牙齿外表面温度逐渐升高。一体式插入器械在牙齿上产生的温度更高,而转接耦合式插入器械在插入器械内部表现出更大的发热。