Kratz Assaf, Yagev Ronit, Belkin Avner, Goldberg Mordechai, Zahavi Alon, Goldberg Ivan, Imtirat Ahed
Department of Ophthalmology, Soroka University Medical Center, Beer-Sheva 84101, Israel.
Faculty of Health Sciences, Ben-Gurion University of the Negev, Beer-Sheva 84101, Israel.
Diagnostics (Basel). 2025 Jul 18;15(14):1813. doi: 10.3390/diagnostics15141813.
: The objective of this study was to evaluate the agreement and clinical interchangeability of intraocular pressure (IOP) measurements obtained with the mechanical Haag-Streit AT900 Goldmann applanation tonometer (mGAT) and the digital Huvitz HT5000 applanation tonometer (dGAT). : This retrospective comparative study included 53 eyes of 28 patients undergoing routine ophthalmologic evaluation. Each eye underwent IOP measurement using both mGAT and dGAT in a randomized sequence. Central corneal thickness (CCT) was also recorded. Pearson's correlation coefficient was used to determine correlation between paired IOP measurements. Bland-Altman plots were graphed for the analysis of differences for IOP between the instruments. : A total of 53 eyes of 28 patients (15 males) were included in the study. The mean age of the patients was 62.6 years. The mean mGAT and dGAT measurements were 16.3 ± 6.6 mmHg (range 9-50) and 16.4 ± 6.2 mmHg (range 8.8-45.9), respectively ( = 0.53). A strong, significant positive correlation was found for paired IOP measurements by the two instruments (r = 0.98; < 0.0001). Bland-Altman analysis revealed 95% limits of agreement from -2.5 to +2.3 mmHg, with a small but statistically significant proportional bias favoring mGAT at higher IOP levels. Additionally, 91% of paired measurements were within ±2 mmHg. CCT-related differences were statistically and clinically insignificant. : IOP measurements obtained with mGAT and dGAT were highly correlated and clinically interchangeable for the range tested. The Huvitz HT5000 may serve as a reliable alternative to the classic Goldmann tonometer in routine clinical settings.
本研究的目的是评估使用机械 Haag-Streit AT900 戈德曼压平眼压计(mGAT)和数字 Huvitz HT5000 压平眼压计(dGAT)测量眼压(IOP)的一致性和临床互换性。本回顾性比较研究纳入了 28 例接受常规眼科检查的患者的 53 只眼。每只眼均以随机顺序使用 mGAT 和 dGAT 测量眼压。同时记录中央角膜厚度(CCT)。采用 Pearson 相关系数确定配对眼压测量值之间的相关性。绘制 Bland-Altman 图以分析两种仪器之间眼压的差异。本研究共纳入 28 例患者(15 例男性)的 53 只眼。患者的平均年龄为 62.6 岁。mGAT 和 dGAT 的平均测量值分别为 16.3±6.6 mmHg(范围 9 - 50)和 16.4±6.2 mmHg(范围 8.8 - 45.9)(P = 0.53)。两种仪器的配对眼压测量值之间存在强的、显著的正相关(r = 0.98;P < 0.0001)。Bland-Altman 分析显示一致性界限为 -2.5 至 +2.3 mmHg,在较高眼压水平存在有利于 mGAT 的小但具有统计学意义的比例偏差。此外,91%的配对测量值在±2 mmHg 范围内。CCT 相关差异在统计学和临床上均无显著意义。在测试范围内,使用 mGAT 和 dGAT 测量的眼压高度相关且在临床上可互换。在常规临床环境中,Huvitz HT5000 可作为经典戈德曼眼压计的可靠替代品。