Kariniemi V, Rovamo L, Pohjavuori M
J Perinat Med. 1985;13(5):233-7. doi: 10.1515/jpme.1985.13.5.233.
The effects of different rejection logic limits on the results of analysis of neonatal heart rate variability from electrocardiograms by a microprocessor-based system were studied on fourteen infants after normal labor and on ten infants after elective cesarean section. In addition to the infant's movements and crying, two further main sources of error in the calculation of variability indices were detected: premature beats and problems in the shape of QRS complexes in a neonatal electrocardiogram. No noteworthy problem was observed in the calculation of the interval index (II), which describes the long term variability of heart rate. In the calculation of the differential indices (DI), which describe the short term component of heart rate, distorted QRS complexes created very high DIs when the rejection logic was ten beats per minute (bpm). When stricter rejection limits were used (five bpm), the DI values even in these cases fell within the normal range. The DI values calculated using a rejection limit of five bpm were always lower than those calculated using a rejection limit of ten bpm, but the reference values of neonatal II and DI after normal labor were similar to those presented previously using a rejection limit of ten bpm. In conclusion, the neonatal electrocardiogram appears to be a noisy signal, comparable with the abdominal fetal electrocardiogram, and strict rejection limits are useful in processing it for indices of variability.
我们对14例顺产婴儿和10例择期剖宫产婴儿进行了研究,以探讨不同的排除逻辑限制对基于微处理器系统分析新生儿心电图心率变异性结果的影响。除了婴儿的运动和哭闹外,在变异性指数计算中还检测到另外两个主要误差来源:早搏和新生儿心电图中QRS波群形态问题。在计算描述心率长期变异性的间期指数(II)时,未观察到明显问题。在计算描述心率短期成分的差值指数(DI)时,当排除逻辑为每分钟10次心跳(bpm)时,形态异常的QRS波群会产生非常高的DI值。当使用更严格的排除限制(5 bpm)时,即使在这些情况下,DI值也落在正常范围内。使用5 bpm排除限制计算的DI值总是低于使用10 bpm排除限制计算的值,但顺产新生儿II和DI的参考值与之前使用10 bpm排除限制时呈现的值相似。总之,新生儿心电图似乎是一个有噪声的信号,与腹部胎儿心电图类似,严格的排除限制有助于对其进行处理以获得变异性指数。