Del Guidice R E, Doering K M, Teran A
Am J Med Technol. 1979 Apr;45(4):287-9.
Researchers evaluated a differential tube solubility test's ability to detect Hemoglobin S (Hb S) in sickle cell conditions. Six hundred twenty-nine blood samples, including 190 S hemoglobinopathies, were used to assess test performance. Alkaline cellulose acetate electrophoresis was used to confirm all results. The test correctly indicated the presence of Hb S in every case in which Hb S was found on electrophoresis. Six of the 198 specimens containing Hb S incorrectly differentiated the heterozygous from the homozygous state (two with Hb AS, three with Hb SF, and one with Hb SC). Electrophoresis later showed four of these samples to be inappropriate for solubility testing. Three of these specimens contained Hb SF and the other specimen contained Hb SC. One specimen from a patient with polycythemia gave a false positive result. Results were compared using different methods of centrifugation. This solubility test was found to be dependable as a screening method provided the manufacturer's directions are followed and all positive results are confirmed by electrophoresis.
研究人员评估了鉴别管溶解度试验在镰状细胞病条件下检测血红蛋白S(Hb S)的能力。使用了629份血样,包括190份S血红蛋白病样本,来评估该试验的性能。采用碱性醋酸纤维素电泳来确认所有结果。该试验在电泳发现Hb S的每一例中都正确地表明了Hb S的存在。198份含有Hb S的标本中有6份错误地区分了杂合子和纯合子状态(2份为Hb AS,3份为Hb SF,1份为Hb SC)。电泳后来显示这些样本中有4份不适合进行溶解度试验。其中3份标本含有Hb SF,另一份标本含有Hb SC。1份来自红细胞增多症患者的标本给出了假阳性结果。使用不同的离心方法比较了结果。发现这种溶解度试验作为一种筛查方法是可靠的,前提是遵循制造商的说明并且所有阳性结果都通过电泳确认。