Falsely high refractometric readings for the specific gravity of pleural fluid.

作者信息

Light R W

出版信息

Chest. 1979 Sep;76(3):300-1. doi: 10.1378/chest.76.3.300.

Abstract

For 128 pleural fluids, the relationship between the protein content and the reading for the specific gravity obtained from a refractometer calibrated for urinary specific gravity was analyzed. The refractometer gave falsely high levels for the specific gravity of the pleural fluid. A reading for 1.019 (rather than 1.016) corresponded to a concentration of protein of 3.0 gm/100 ml, and each deviation of 0.005 (rather than 0.003) coresponded to a concentration of 1 gm/100 ml; however, determinations of the concentration of protein in the pleural fluid directly from the refractometer's scale for protein (calibrated for serum) was rapid and accurate. Calculation of the protein content of the pleural fluid from the reading for specific gravity on the refractometer is erroneous and sometimes misleading.

摘要

对128份胸腔积液样本,分析了蛋白质含量与使用针对尿比重校准的折射仪测得的比重读数之间的关系。该折射仪给出的胸腔积液比重读数偏高。读数为1.019(而非1.016)时对应的蛋白质浓度为3.0克/100毫升,每偏差0.005(而非0.003)对应的蛋白质浓度为1克/100毫升;然而,直接根据折射仪针对蛋白质的刻度(针对血清校准)测定胸腔积液中的蛋白质浓度既快速又准确。根据折射仪上的比重读数计算胸腔积液的蛋白质含量是错误的,有时还会产生误导。