Suppr超能文献

Determination of the dew point and the frost point below 0 degrees C making use of the beta-ray backscattering and the electric conductivity on the narrow surface of insulated layer.

作者信息

Matsumoto S, Kobayashi H

出版信息

Radioisotopes. 1979 Oct 15;28(10):609-12. doi: 10.3769/radioisotopes.28.10_609.

Abstract

It is necessary to distinguish between the dew point and the frost point below 0 degrees C. The freezing of the dew and the melting of the frost are respectively detected by the rapid decrease and the increase of the conduction current on the narrow surface of insulated layer made of epoxy, 0.5 mm in width and 10 mm in length, on which the dew deposits. The dew point -9 degrees C and the frost point -8 degrees C in the humidity 21% at the temperature 13 degrees C are clearly distinguished in this method.