Nelson R M, Soderberg G L
Electroencephalogr Clin Neurophysiol. 1983 Feb;55(2):238-9. doi: 10.1016/0013-4694(83)90194-3.
To selectively record a limited number of skeletal muscle motor unit action potentials, a bifilar fine wire electrode was developed. Two 8 micrometers wires were braided together, a hook formed, and ends insulated. A round beam, pulsed laser, etched a 5 micrometers piece of insulation from each wire near the hooked end. Bipolar recordings from a select number of skeletal muscle motor units were obtained from these two uninsulated 5 micrometers areas. Reactive and capacitative components for each set of electrodes were measured. These electrodes were selective in recording from 3 to 5 single motor unit action potentials at low isometric tension levels.
为了选择性记录有限数量的骨骼肌运动单位动作电位,开发了一种双线细丝电极。将两根8微米的导线编织在一起,形成一个钩子,并将末端绝缘。使用圆形光束脉冲激光在靠近钩状端的每根导线上蚀刻掉5微米的绝缘层。从这两个5微米的非绝缘区域获得了选定数量的骨骼肌运动单位的双极记录。测量了每组电极的电抗和电容成分。这些电极在低等长张力水平下能够选择性地记录3至5个单个运动单位动作电位。