Orthogonal ventricular electrogram sensing.

作者信息

Goldreyer B N, Brueske R, Knudson M B, Cannom D S, Wyman M G

出版信息

Pacing Clin Electrophysiol. 1983 Jul;6(4):761-8. doi: 10.1111/j.1540-8159.1983.tb05337.x.

Abstract

Inappropriate demand pacing is most commonly due to improper ventricular electrogram sensing. Filters and programmable sensitivities improve electrogram sensing of conducted beats, but paced electrograms cannot be sensed by conventional unipolar or bipolar systems. A permanent pacing lead with a standard tip electrode and three orthogonal 0.8 mm2 sensing electrodes located circumferentially 2 cm proximal to the pacing tip was tested in 22 patients. The tip electrode was placed in the right ventricular apex in standard pacing position. Orthogonal electrodes were not in contact with ventricular myocardium. Orthogonal ventricular electrograms from 54 electrode pairs were compared with unipolar tip electrograms during conducted rhythms and paced beats. Tip ventricular electrograms averaged 12.8 mV with 3.04 mVT waves. Orthogonally recorded ventricular electrograms during conducted beats averaged 8.86 mV with T waves of 1.57 mV. During pacing, tip ventricular electrograms were obscured by the stimulus artifact and repolarization events. Orthogonal ventricular electrograms, however, demonstrated small discrete stimuli of 1.99 mV followed by discrete ventricular electrograms of 9.19 mV and T waves of 1.9 mV. Orthogonal ventricular electrograms compared favorably with contacting tip electrograms during conducted beats and provided a redundant sensing capability. During pacing, orthogonal ventricular electrograms allowed the capability for capture verification. A new pacing catheter allows for improved ventricular electrogram sensing and capture verification.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验