Yuen B H, Pride S M, Callegari P B
Can Med Assoc J. 1984 Sep 15;131(6):583-5.
We tested the validity of the beta-Neocept test for early pregnancy against that of the plasma human chorionic gonadotropin beta-subunit radioimmunoassay (beta-HCG RIA). The beta-Neocept test had a sensitivity of 88%, a specificity of 93%, a positive predictive value of 95%, a negative predictive value of 84% and an accuracy of 90%. In view of these performance characteristics, its low cost and its ease of use, the beta-Neocept test could be used as the initial pregnancy test when there is a high probability of pregnancy, as there was in this study population, which consisted of 111 women attending endocrine infertility clinics. The more expensive beta-HCG RIA could be reserved for special indications and for patients in whom the results of the urinary hemagglutination inhibition tests are inconsistent with the clinical signs and symptoms.
我们将β-新孕测试用于早期妊娠的有效性与血浆人绒毛膜促性腺激素β亚基放射免疫测定法(β-HCG RIA)进行了比较。β-新孕测试的灵敏度为88%,特异性为93%,阳性预测值为95%,阴性预测值为84%,准确率为90%。鉴于这些性能特征、低成本以及易于使用,当怀孕可能性较高时,如本研究人群(由111名前往内分泌不孕症诊所就诊的女性组成)的情况,β-新孕测试可作为初始妊娠测试。更昂贵的β-HCG RIA可保留用于特殊指征以及尿血凝抑制试验结果与临床体征和症状不一致的患者。