Buchel L A, Delain E, Bouteille M
J Microsc. 1978 Mar;112(2):223-9. doi: 10.1111/j.1365-2818.1978.tb01168.x.
The fluorographic process for enhancement of the autoradiographic detection of beta-ray emitted by tritium has been successfully adapted to electron microscope autoradiography. An original scintillator plastic film helped to increase the detection yield of electrons up to 80% without alteration of the morphological aspect of the autoradiograms, and without increasing background fog or causing apparent loss of resolution.
用于增强氚发射的β射线放射自显影检测的荧光成像方法已成功应用于电子显微镜放射自显影。一种原始的闪烁体塑料薄膜有助于将电子检测率提高到80%,而不会改变放射自显影片的形态,也不会增加背景雾度或导致明显的分辨率损失。