Levine P A, Caplan C H, Klein M D, Brodsky S J, Ryan T J
Chest. 1982 Oct;82(4):461-5. doi: 10.1378/chest.82.4.461.
The effect of isometric upper extremity exercise on pacemaker function was evaluated in 27 patients who remained pacemaker-dependent during testing. Inhibition was demonstrated in eight (31 percent) of which five were symptomatic. Based on design of the sensing amplifier and return to an all-metal housing in the unipolar lithium pulse generators, myopotential inhibition is being recognized as one cause of symptomatic pacemaker inhibition that is more common than generally appreciated. A method of evaluation and management options for symptomatic patients are discussed. Routine testing of all patients should be performed at the time of a regular office evaluation. If one model pulse generator appears to be particularly prone to myopotential inhibition, this knowledge should be considered in the choice of future pacing systems.
在27例测试期间仍依赖起搏器的患者中评估了等长上肢运动对起搏器功能的影响。其中8例(31%)出现抑制,5例有症状。基于传感放大器的设计以及单极锂脉冲发生器采用全金属外壳,肌电位抑制被认为是有症状起搏器抑制的一个原因,其比通常认为的更为常见。本文讨论了有症状患者的评估方法和管理选择。所有患者应在定期门诊评估时进行常规测试。如果某一型号的脉冲发生器似乎特别容易出现肌电位抑制,那么在选择未来的起搏系统时应考虑这一情况。