Monbaliu J G, Rosseel M T, Bogaert M G
J Pharm Sci. 1981 Aug;70(8):965-6. doi: 10.1002/jps.2600700840.
A high-performance liquid chromatographic method for the determination of methoxsalen in plasma was developed. The method includes extraction from plasma of the drug and the internal standard (5-methoxypsoralen) into methylene chloride. Chromatography was performed on a reversed-phase C8 column connected with a UV detector set at 254 nm. The mobile phase was methanol-acetonitrile-water (2:30:68). For a plasma sample of 0.25 ml, the maximal sensitivity was approximately 10 ng/ml. Accuracy was within 7.5% for therapeutic plasma levels, and the coefficient of variation varied between 4.3 and 0.9% for 28 and 300 ng/ml of plasma, respectively.
建立了一种测定血浆中补骨脂素的高效液相色谱法。该方法包括将药物和内标(5-甲氧基补骨脂素)从血浆中萃取到二氯甲烷中。色谱分析在与设定于254nm的紫外检测器相连的反相C8柱上进行。流动相为甲醇-乙腈-水(2:30:68)。对于0.25ml的血浆样品,最大灵敏度约为10ng/ml。治疗血浆水平的准确度在7.5%以内,血浆浓度为28ng/ml和300ng/ml时变异系数分别在4.3%和0.9%之间。