Unger H, Willms B
Dtsch Med Wochenschr. 1980 Apr 18;105(16):566-70. doi: 10.1055/s-2008-1070707.
Blood sugar self-control was assessed in 20 in-patients using a new blood sugar test strip, Haemo-Glukotest 20--800. Comparative estimations of 445 strip tests with values obtained by the hexokinase method showed good correlations particularly in the low range below the renal threshold. Nearly 70% of measurements were assessed correctly in this range. Mean deviations of test strip results from the corresponding laboratory value were 0.61 to 0.89 mmol/l (11--16 mg/dl). Haemo-Glukotest 20--800 is thus well suited for blood glucose self-control particularly in the range below the renal threshold. In a survey most patients estimated the importance of blood sugar to be more than that of urinary sugar.
使用新型血糖测试条Haemo-Glukotest 20-800对20名住院患者的血糖自我控制情况进行了评估。将445次测试条检测结果与通过己糖激酶法获得的值进行比较估计,结果显示二者具有良好的相关性,尤其是在低于肾阈值的低范围内。在此范围内,近70%的测量结果评估正确。测试条结果与相应实验室值的平均偏差为0.61至0.89毫摩尔/升(11-16毫克/分升)。因此,Haemo-Glukotest 20-800非常适合血糖自我控制,尤其是在低于肾阈值的范围内。在一项调查中,大多数患者认为血糖的重要性高于尿糖。