Weidlich E, Richter G J, Mund K, von Sturm F, David E, Brandt G
Med Prog Technol. 1980 Apr;7(1):11-9.
In the skeletal muscle of cats, semispherical stimulating electrodes having a radius of 1.2 mm deliver a mean threshold current of 0.15 mA, corresponding to 1.7 mA/cm2. These values are doubled within a period of four weeks after chronic implantation. In the case of non-polarizable electrodes the mean threshold voltage increases from 108 mV to 171 mV. The lowest individual values range between 50 mV and 60 mV at the time of implantation. The highest value is around 500 mV after four weeks. The high values have been observed with strongly polarizing ELGILOY electrodes. The chronic threshold energies vary between 3 x 10(-8) Ws for activated vitreous carbon electrodes and 17.5 x 10(-8) Ws for ELGILOY electrodes. There are two reasons for the low threshold energy of activated vitreous carbon electrodes. One reason is their low polarization, and the other is their better compatibility. Connective tissue layers with a thickness between 25 micron and 50 micron are found around well healed carbon electrodes.
在猫的骨骼肌中,半径为1.2毫米的半球形刺激电极的平均阈值电流为0.15毫安,相当于1.7毫安/平方厘米。在长期植入后的四周内,这些值会翻倍。对于非极化电极,平均阈值电压从108毫伏增加到171毫伏。植入时最低的个体值在50毫伏至60毫伏之间。四周后最高值约为500毫伏。在强极化的埃尔吉洛伊耐蚀合金电极上观察到了高值。慢性阈值能量在活性玻璃碳电极的3×10⁻⁸瓦秒到埃尔吉洛伊耐蚀合金电极的17.5×10⁻⁸瓦秒之间变化。活性玻璃碳电极阈值能量低有两个原因。一个原因是它们的低极化,另一个原因是它们更好的兼容性。在愈合良好的碳电极周围发现了厚度在25微米至50微米之间的结缔组织层。