[视神经超声成像的新知识。实验与临床研究]

[New knowledge of echographic imaging of the optic nerve. Experimental and clinical studies].

作者信息

Hasenfratz G, DeLa Torre M D, Haigis W

机构信息

Universitäts-Augenklinik Würzburg.

出版信息

Ophthalmologe. 1995 Feb;92(1):51-5.

PMID:7719077
Abstract

The optic nerve can be displayed in vivo by A- and B-scan techniques. To achieve correlations between the anatomical structure of the optic nerve and the echograms and to answer the question if by the usual examination techniques the diameter of the nerve can be measured, experimental studies on bovine optic nerves were carried out: examination of the bovine optic nerve in vitro with the A-scan technique; measurement of the sound velocity in bovine optic nerves. Additionally, we compared the two most common A-scan techniques for the display of the optic nerve in vivo in 47 normal persons and in 35 patients with changes of the optic nerve. The experimental studies showed that the orbital fat and outer sheaths of the optic nerve cause refraction of the sound beam. The sound velocity of ultrasound in bovine optic nerves was measured to 1567 +/- 25 m/s. The two most common examination techniques in vivo showed no differences in normal persons, but significant differences in patients with increased subarachnoidal fluid.

摘要

通过A扫描和B扫描技术可在活体中显示视神经。为了实现视神经解剖结构与超声图像之间的关联,并回答能否通过常规检查技术测量神经直径这一问题,我们对牛视神经进行了实验研究:用A扫描技术在体外检查牛视神经;测量牛视神经中的声速。此外,我们在47名正常人及35名视神经有病变的患者中比较了两种最常用的A扫描技术在活体中显示视神经的情况。实验研究表明,眼眶脂肪和视神经外鞘会导致声束折射。测得超声在牛视神经中的声速为1567±25米/秒。两种最常用的活体检查技术在正常人中无差异,但在蛛网膜下腔积液增多的患者中有显著差异。