Harmon R L, Rodriquez A A
Department of Rehabilitation Medicine, University of Wisconsin-Madison Medical School.
Electromyogr Clin Neurophysiol. 1994 Jul-Aug;34(5):275-8.
Using a combined concentric (CN) and modified single fiber (MSF) electromyographic needle electrode with both recording surfaces at the tip approximately 25 microns apart, 24 motor unit potentials (MUPs) were recorded from one extensor digitorum communis muscle, triggering from either the CN or MSF channel in equal numbers and averaging both CN and MSF recordings. Filter settings were 20-10,000 Hz. Correlations of MUP amplitude, phases, turns, negative turns (NTs), and main spike duration, but not rise time or duration, between the CN and MSF recordings were statistically significant (P < 0.05). Mean +/- SD numbers of estimated single fiber potentials (ESFPs) and NTs for the MSF recordings were 1.17 +/- 0.38 and 1.54 +/- 0.72. Correlations between numbers of ESFPs and turns, NTs and phases, NTs and turns, and NT and main spike duration for the MSF and CN MUP recordings, respectively, were statistically significant (P < 0.05). The study findings using these techniques continue to support evidence that one or a few muscle fibers contribute to the main spike component of the MUP.
使用一种组合式同心(CN)和改良单纤维(MSF)肌电图针电极,其两个记录表面位于针尖处,相距约25微米,从一块指总伸肌记录到24个运动单位电位(MUP),分别从CN或MSF通道触发,数量相等,并对CN和MSF记录进行平均。滤波设置为20 - 10,000赫兹。CN和MSF记录之间的MUP振幅、相位、转折、负向转折(NT)和主波峰持续时间的相关性具有统计学意义(P < 0.05),但上升时间或持续时间的相关性无统计学意义。MSF记录的估计单纤维电位(ESFP)和NT的平均±标准差数量分别为1.17±0.38和1.54±0.72。MSF和CN MUP记录中ESFP数量与转折、NT与相位、NT与转折以及NT与主波峰持续时间之间的相关性分别具有统计学意义(P < 0.05)。使用这些技术的研究结果继续支持这样的证据,即一根或几根肌纤维对MUP的主波峰成分有贡献。